September 1973 U.S. ATOMIC ENERGY COMMISSION
REGULATORY GUIDE
DIRECTORATE OF REGULATORY STANDARDS
REGULATORY GUIDE 4.3 MEASUREMENTS OF RADIONUCLIDES IN THE ENVIRONMENT
ANALYSIS OF 1-131 IN MILK
A. INTRODUCTION
"Monitoring Radioactivity Releases," of Appendix A to 10 CFR Part 50, "General Design Criteria for Nuclear Power Plants,"
requires that means be provided for monitoring the plant environs of light-water-cooled nuclear power plants for radioactivity that may be released from normal operations, including anticipated operational occurrences, and from postulated accidents.
Paragraph (a)(2) of § 50.36a of 10 CFR Part 50
provides that technical specifications for each license will include a requirement that the licensee submit a report to the Commission within 60 days after January 1 and July 1 of each year which, in addition to specifying S
the quantity of each of the principal radionuclides released to unrestricted areas in liquid and airborne effluents during the last six months of operation, provides sufficient information to estimate annual radiation doses to the public resulting from effluent releases.
Paragraph (e) of § 20.106 of 10 CFR Part 20
provides that the Commission may limit the quantities of radioactive materials released in air or water by licensees during a specified period of time if it appears that the daily intake of radioactive materials from air, water, or food by a suitable sample of an exposed population group, averaged over a time period not exceeding one year, would otherwise exceed specified quantities. Section 20.201 of 10 CFR 20 requires that a licensee conduct surveys of levels of radiation or concentrations of radioactive material as necessary for compliance with AEC regulations in Part 20. Paragraph (c) of § 20.1 of 10 CFR Part 20 states that every reasonable effort should be made by AEC licensees to maintain radiation exposures, and releases of radioactive materials in effluents to unrestricted areas, as far below the limits specified in Part 20 as practicable, i.e., as low as is practicably achievable, taking into account the state of technology, and the economics of improvements in relation to benefits to the public health and safety and in relation to the utilization of atomic energy in the public interest.
The procedure for the determination of 1-131 in milk described in this guide constitutes a method acceptable to the Regulatory staff for measuring 1-131 as may be required by the above regulations or orders or license conditions imposed pursuant thereto.
B. DISCUSSION
In accord with the Commission's policy of keeping exposure of man to radioactive materials released from licensed activities, including the operation of production or utilization facilities, as low as practicable, licensees should make every reasonable effort to limit the release of radioactive materials in effluents to unrestricted areas as far below the limits specified by the Commission as practicable.
Iodine-131 requires special attention because it is one of the more significant radionuclides present in releases from nuclear power plants, and it can be concentrated through the air-vegetation-cow-milk pathway and potentially contribute to the thyroid dose.
Although most milk data have indicated little or no
1-131, milk monitoring at the suggested improved level of sensitivity would provide direct evidence that the potential exposure due to
1-131 does not exceed predicted values.
Current guidelines' for 1-131 dose to the thyroid of infants require that the sensitivity for measuring 1-131 in milk be improved over conventional direct gamma counting of the raw milk sample.
Low-level beta counting techniques with a detection sensitivity about an order of magnitude better than conventional gamma counting techniques can be used to attain the desired sensitivity. In the analysis of 1-131 in milk, it may be
'Regulatory Guide 1.42, "Interim Licensing Policy on As Low As Practicable for Gaseous Radioiodine Releases from Light-Water-Cooled Nuclear Power Reactors."
USAEC REGULATORY GUIDES
Copies of published guides may be obtained by request indicating the divisions desired to the U.S.
Atomic Energy Commission, Washington, D.C. 20545, Regulatory Guides are issued to describe and make available to the public Attention: Director of Regulatory Standards. Comments and suggestions for methods acceptable to the AEC Regulatory staff of implementing specific parts of improvements in these guides are encouraged and should be sent to the Secretary the Commission's regulations, to delineate techniques used by the staff in of the Commission, U.S. Atomic Energy Commission, Washington, D.C. 20545, evaluating specific problems or postulated accidents, or to provide guidance to Attention: Chief, Public Proceedings Staff.
applicants. Regulatory Guides are not substitutes for regulations and compliance with them is not required. Methods andsolutions different from those set out in The guides are issued in the following ten broad divisions:
the guides will be acceptable if they provide a basis for the findings requisite to the issuance or continuance of a permit or license by the Commission.
1. Power Reactors
6. Products
2. Research and Test Reactors
7. Transportation
3. Fuels and Materials Facilities B. Occupational Health Published guides will be revised periodically, as appropriate, to accommodate
4. Environmental and Siting
9. Antitrust Review comments and to reflect new information or experience.
5. Materials and Plant Protection
1
0. General
necessary in some instances to measure 1-131 in the presence of 1-1292 as well as other radioisotopes of iodine; since the usual beta counting methods are nonspecific, provisions for the resolution of such a potential mixture of isotopes should be included in the procedure. In most instances, however, radioisotopes of iodine other than 1-131 and possibly 1-129 are not expected to be of significance in the milk pathway because of their short half-lives. Also, under normal operating conditions of nuclear power plants, the contribution from 1-129 will be negligible, the potential significance of 1-129 being associated primarily with fuel reprocessing plants. It is probable, therefore, that any radioiodine detected in milk sampled from the environs of nuclear power plants will be primarily
1-131.
Nevertheless, because of the nonspecificity of beta counting methods, as a minimum, decay checks should be performed.
A simple check for isotopic purity can be performed by recounting the sample for radioactive decay. A
half-time of 8 days would be a good indication that the activity is due to 1-131. However, since it is not practical to attempt to monitor decay half-times when the sample net counting rate is less than about 0.3 cpm, it would be
2A regulatory guide for the analysis of 1-129 in milk is in preparation.
worthwhile to do this only when the net count is more than about 0.3 cpm. If, after the second count, it is determined that the activity is not decaying with a half-time of about 8 days, a third count made 4-5 days after the second count can provide the basis for a rough graphical resolution of 1-131 activity. In any case, it would be desirable to hold the sample for at least a year to permit possible identification of the long-lived component should the need arise.
C. REGULATORY POSITION
The procedure for the analysis of 1-131 in milk described in the appendix to this guide is acceptable to the Regulatory staff as a basis for meeting analytical performance standards which are compatible with established guidelines.'
Other procedures selected for analyzing
1-131 in milk should provide levels of analytical sensitivity and specificity similar to those for the method described in the appendix.
See footnote on page 4.3-1
4.3-2
APPENDIX
IODINE-131 ANALYSIS OF MILK AT CONCENTRATIONS
DOWN TO 0.25 PICOCURIE PER LITER
The direct ion exchange method for separating iodine from raw milk requires that the iodine be in a readily exchangeable anionic form. Radioiodine tracer experiments on cows' 2 3 have shown that 0-10 percent of the iodine in milk may be protein bound, rendering this fraction essentially unavailable for exchange with the ion exchange resin. However, this uncertainty of up to
10 percent in the 1-131 determination will not significantly affect data interpretation and therefore is considered to be acceptable. Nevertheless, for the sake of accuracy, it is recommended that a fixed correction factor of 1.05(1/0.95 = 1.05) be applied to the counting data to compensate for this effect.
Also, there may be instances where milk samples may curdle to varying degrees in the interim between collection and analysis. Milk in this condition is unsuitable for analysis by the direct ion exchange method. A fresh sample should be obtained should this occur. If clabbering continues to be a problem, the ion exchange separation should be performed at the sample collection point and the ion exchange resin column shipped to the laboratory for, processing. If this cannot be done, the milk sample should be frozen prior to shipment to the analytical laboratory.
The procedure described in this appendix was adapted from a procedure entitled "Determination of
1-129 in Milk and Water Samples" developed by the Radiological Sciences Laboratory, Division of Laboratories and Research, New York State Department of Health. It was tested at the Battelle Memorial Institute Pacific Northwest Laboratories and shown to be adequate to measure 1-131 in milk at the 0.25 picocurie per liter concentration level. This sensitivity is based on using a 4-liter milk sample and beta counting for 1000 minutes or longer in a low-background counter with a nominal background count rate of 0.5-1 count per minute (cpm). The analytical sensitivity can be further improved by using more than 4 liters of milk and counting for longer than 1000 minutes.
A recount for decay check should be performed
7-10 days after the first count. However, because it is not practical to do this for net counting rates of less than about 0.3 cpm, a recount should be made only when the initial net counting rate is greater than 0.3 cpmn. If, after the second count, it is determined that the activity is not decaying with a half-life of 8 days, a third count should be made 4-5 days after the second count.
'Murthy. G. K., et al "Method for Removing Iodine-131 from Milk," Journal of Dairy Science. 45:1066 (1962)
2Saiki, M.. Omomo, Y., "Determination of Iodine-1 31 in Milk, Quick Methods for Radiochemical Analysis," Technical Reports Series No. 95, IAEA, 1969.
' Bretthauer, E. W., Mullen. A. L., Moghissi, A. A. "Milk Transfer Comparisons of Different Chemical Forms of Radioiodine," Health Physics, 22:257 (1972)
A graphical resolution of 1-131 activity based on these three points may be performed. The sample should be held for at least a year to permit possible identification of the long-lived component should the need arise.
Abstract of the Procedure Iodine-131 is removed from milk by concentrating the iodine on an anion exchange resin column and subsequently removing it from the resin by batch extraction using NaOCI.
After reduction to
12 by hydroxylamine hydrochloride, the iodine is extracted into CC14 , reduced with bisulfite, and back extracted into water. The iodine is precipitated as palladous iodide. Chemical, yield based on the added carrier iodine is determined gravimetrically. The 1-131 concentration is determined by counting the palladous iodide precipitate in a low-background beta counter. The yield for the procedure is 70 to 85%.
Special Apparatus
1. Ion exchange column, 2 cm dia. by about 10.cm long with standard taper fittings
2. Four-liter reservoir with standard taper fitting
3. Assorted sizes of glass beakers
4. Hot plate
5. Magnetic stirrer
6. Suction filtering apparatus
7. 250-ml and 125-ml separatory funnels
8. Clinical centrifuge
9. Low-background beta counter (nominal background
<1 count per min.)
10. Nylon rings and discs, Mylar, glass fiber paper (2.8 cm) (see HASL 300)
Reagents and Chemicals
1. Iodide carrier, 10 mg/ml - dissolve 1.181 grams Nal in 100 ml water
2. Dowex I x 8, 20-50 mesh, Cl" form
3. 2M NaCI - dissolve 116.9 grams NaCI in I liter water
4. NaOCI, 5-6% (commercial strength)
5. HNO 3 , concentrated
6. Hydroxylamine hydrochloride, crystals
7. CCI4
8. IM NaHS0 3 - dissolve 1.041 grams of NaHSO 3 in
10 ml water (prepare daily)
9. Palladous chloride, 10 mg/ml - dissolve 1.66 grams ofPdCl2 in 100 ml of water
4.0-3
Procedure
1. Prepare ail ion exchange column 2 cm in diameter by 10 cm long and fill to a height of 5 cm with Dowex 1 x 8, 20-50 mesh, Cl- form. Add the resin from a water slurry.
2. Add 1-2 ml of iodine carrier (10 mg '/ml) to 4-liter milk sample and stir thoroughly. Accurately measure and record exact volume of carrier added.
Pass through the column at a flow rate of about 30
ml per minute and discard effluent.
3. Wash the column with 500 ml hot (500C) distilled water followed by 100 ml of 2M NaCI. Discard washes.
4. Transfer the resin to a 250-mil beaker using 50 ml of
5-6% NaOCI. (Note 1)
5. Place a small (-2 cm long) magnetic stirring bar in the beaker and stir vigorously for 5 minutes on a magnetic stirrer.
6. Filter the resin slurry through a suction filter and retain the NaOCI solution, (Note 2)
7. Reextract the resin by repeating steps 4, 5, and 6.
8. Discard the resin, combine the two 50-ml solutions, and carefully add 20 ml of conc. HNO 3. (Note 3)
9. Pour the acidified NaOCI solution into a 250-ml separatory funnel and add 50 ml of CC14 .
10. Add 1.5 gin of hydroxylamine hydrochloride and shake (Note 4). Extract the iodine into the organic phase (about 2 min. equilibration). (Note 5)
11. Drain lower organic phase into a clean 250-nl separatory funnel and save.
1
2. Add
50
ml CC14 and I gm hydroxylamine hydrochloride to the aqueous phase in the first separatory funnel and reextract. Combine organic phases and discard aqueous phase.
I
13. Add 25 ml H20 and 10 drops of freshly prepared IM NaHSO 3 to the separatory funnel containing the combined CC14 and shake.
Equilibrate for 2 minutes. Discard organic (lower) phase. (Note 6)
14. Transfer the aqueous (upper) phase into a clean
50-ml centrifuge tube and add I ml of conc. HNO 3 and 10 ml of PdCI2 solution. Stir and let stand for 5 minutes.
15. Centrifuge Pd12 precipitate discarding supernate.
Wash precipitate by stirring with 10 ml of H20.
16. Using a filter funnel setup similar to that (Teflon or polyethylene - nylon) described in HASL 300,
"Procedure Manual," USAEC, filter with suction through a tared glass fiber paper (2.8 cm diameter),
using a water wash bottle to effect the transfer.
(Note 7)
17. Dry precipitate for 20 minutes in an oven set at I 100C and weigh to the nearest 0.1 milligram.
(Alternatively, the sample may be dried on a hot plate at low heat for about 30 minutes.)
18. Mount the precipitate on a nylon disc, cover with Mylar (1/4 nril thick), and fasten with ring. (This procedure is described in HASL 300.)
19. Count in a low-background counter for 1000
minutes.
20. If net counting rate of sample is greater than 0.3 cpm, recount after 7-8 days.
21. Calculate as picocuries 1-131 per liter of milk at time of sampling.
Notes
1. This is most easily accomplished by using a polyethylene squirt bottle to dispense the NaOCI.
2. Resin should be very light straw color after NaOCI
extraction; if not light colored, the NaOCI is below strength, and a fresh solution should be obtained.
Commercial grade Clorox has been found to be adequate.
3. Add the acid slowly with stirring until the vigorous reaction subsides. Perform in well ventilated hood.
4. Proceed with caution in this step. Excessive gas formation during the extraction can cause the stopcock or cap on the separatory funnel to "pop"
with consequent loss of sample. Start by gently swirling the solution to effect mixing. Invert the separatory funnel with the stopcock pointing tip and release the pressure by opening the stopcock.
Close the stopcock, shake, and repeat the pressure release sequence.
5. Organic phase should be deep red; if not, add additional gram of NH 2 OH-HCI.
6. After back extraction into water, CC14 should be colorless; if not add additional NaHSO 3 and reextract.
7. HASL 300 "Procedure Manual" Health and Safety Laboratory, U.S. Atomic Energy Commission, 376 Hudson St., New York, New York 10014.
Calibration and Standardization of Beta Counter Counter standardization should be performed with 1-131 standard obtained from the National Bureau of Standards (NBS) or with one which is relatable to NBS.
An example of a relatable standard is a secondary or tertiary standard which has been calibrated against a primary NBS standard.
A
self-absorption calibration curve should be constructed using at least 4 points. A useful curve can be constructed by plotting counter efficiency against mass.
The counter efficiency can then be read off the curve as a function of the gravimnetric yield.
4.3-4
- .
Example of Calculation of Results The activity in picocuries per liter at the time the milk was collected is determined by calculations based on the observed net counts per minute of 1-131. The calculation requires correction for chemical yield, protein-bound iodine, counter efficiency, decay since sampling, and the sample volume.
An example of a calculation of the 1-131 concentration in a sample and the error (standard deviation) associated with the measurement follows:
Wt of ppt, ing Chemical yield =
14.83 *
- The ppt is not stoichiometrically Pdl 2 , and this value is a determined empirical yield for 100% recovery; 14.83 mg is the determined weight of Pdl 2 for 10 mgof V.
I.
Chemical yield
2.
Counter efficiency
3.
Counter background
4.
Counting time
5.
Radioactive decay
6.
Sample net counts Gross counts for 1000-min interval The net counts Standard deviation Net count with error limit Net pCi/liter
=
0.78
=
0.35
=
0.6 cpm
1000 minutes
8.05 days
=
0.6 cpm (for a 1-half-life decay period)
(background cpin + net cpm) (counting time)
=
(0.6 + 0.6) 1000 = 1200
=
1200 - 600 = 600 counts per 1000 minutes or 0.6 cpm
=V gross counts + background counts If 1200 +600 = 42
=
600 +/- 42 (net counts) (factor for protein-bound fraction)
(count time)( chem yield)( counter eff)( sample vol)(decay)(dpm/pCi)
(600 +/- 42) (1.05)
('1000)(0.78.)(0.35.)(4)('0.50' 1 2.)
0.
w
=
0.5 pCi/liter with standard deviation of +- 7%
4.3-5
May 1974 U.S. ATOMIC ENERGY COMMISSION
REGULATORY
DIRECTORATE OF REGULATORY STANDARDS
GUIDE
REGULATORY GUIDE' 4.6 MEASUREMENTS OF RADIONUCLIDES IN THE ENVIRONMENT
STRONTIUM-89 AND STRONTIUM-90 ANALYSES
A. INTRODUCTION
General Design Criterion 64,. "Monitoring Radioactivity Releases,"
of Appendix A, "General Design Criteria for Nuclear Power Plants," to. 10 CFR.
Part 50, "Licensing of Production and Utilization Facilities,"
requires that means be provided for monitoring the plant environs of light-water-cooled nuclear power plants for radioactivity that may be released from normal operations, including anticipated operational occurrences, and from postulated accidents.
Paragraph 50.36a (a)(2) of 10 CFR Part 50 provides that technical specifications for each operating license for a nuclear power reactor will include a requirement that the licensee submit a report to the Commission within 60 days after January 1 and July 1 of each year which, in addition to specifying the quantity of each of the principal radionuclides released to unrestricted areas.
in liquid and airborne effluents during the previous six months of operation, provides sufficient information to estimate annual radiation doses to the public resulting from effluent releases.
Paragraph 20.106(e) of 10 CFR Part 20,"Standards for Protection Against Radiation," provides that the Commission may limit the quantities of radioactive materials released in air or water by licensees during a specified period of time if it appears that the daily intake of radioactive materials from air, water, or food by a suitable sample of an exposed population group, averaged over a time period not exceeding one year,
ýwould otherwise exceed specified quantitie
s. Section
20.201, "Surveys," of 10 CFR Part 20 requires that a licensee conduct surveys of levels of radiation or concentrations of radioactive material as necessary for compliance with AEC regulations in 10 CFR Part 20.
Paragraph 20.1(c) of 10 CFR Part 20 states that every reasonable effort should be made by AEC licensees to maintain radiation exposures, and releases of radioactive materials in effluents to unrestricted areas, as far below the limits specified in Part 20 as practicable, i.e., as low as is practicably achievable, taking into account the state of technology and. the economics of improvements in relation to benefits to the public health and safety and in relation to the utilization of atomic energy in the public interest.
This guide describes radioanalytical procedures acceptable to the Regulatory staff' for implementing these requirements with respect to the determination of Sr-89 and Sr-90 in the environment.
B.: DISCUSSION
Radioisotopes of strontium, principally Sr-90 and Sr.89, are fission products, of concern because they are chemically similar to calcium and deposit in bone. The literature lists a large number, of procedures for conducting Sr-90 and Sr-89 analyses, and a wide range of analytical sensitivities maybe obtained depending on the procedure . used; no single procedure, however, is applicable to all, situations. This guide is intended to provide a selection of radioanalytical procedures which have the required performance standards of sensitivity and accuracy for the determination of Sr-89 and Sr-90 in various media.
The levels of Sr-90 and Sr-89 in environmental media are normally quite low, Sr-89 being measurable only for relatively short periods after its release to the environment. Large sample sizes and low-background beta counters are usually required in analyzing for Sr-90
and Sr-89. Although procedures have been developed for analyzing a medium such as milk without prior ashing or concentration of the sample, most methods for analyzing biological and environmental samples require some degree of sample preparation prior to analysis. This USAEC REGULATORY GUIDES
Copies of published guides may be obtained by request indicating the divisions desired to the US. Atomic Energy Commission, Washington, D.C. 20545, Regulatory Guides are issued to describe and make available to the public Attention: Director of Regulatory Standards. Comments and suggestions for methods acceptable to the AEC Regulatory staff of implementing specific parts of Improvements in these guides are encouraged and should be sent to the Secretary the Commission's regulations, to delineate techniques used by the staff in of the Commission, US. Atomic Energy Commission, Washington, D.C. 20545, evaluating specific problems or postulated accidents, or to provide guidance to Attention: Chief, Public Proceedings Staff.
applicants. Regulatory Guides are not substitutes for regulations and compliance with -them is not required. Methods and solutions different from thou set out in The guides are Issued in the following ten broad divisions:
the guides will be acceptable if they provide a basis for the findings requisite to the issuance or continuance of a permit or license by the Commission.
1. Power Reactors
6. Products
2. Research and Test Reactors
7. Transportation
3. Fuels and Materials Facilities
8. Occupational Health Published guides will be revised periodically, as appropriate, to accommodate
4. Environmental and Siting
9. Antitrust Review comments and to reflect new information or experience.
5. Materials and Plant Protection
1
0. General
usually implies wet or dry ashing of the sample. If Sr-90
alone is to be determined, the use of Sr-85 tracer for radiochemical yield determination is recommended, but if both Sr-90 and Sr-89 analyses are required, Sr-85 tracer should not be used because it interferes with the Sr-89 measurement. It is necessary in the latter case to do the yield determination by a gravimetric technique.
One of the most important steps in the analysis is the early exchange of strontium carrier and/or tracer with the Sr-89 and Sr-90 in the sample. Equilibration can be considered to have been attained when the radiostrontium in the sample has been completely solubilized in the presence of strontium carrier and/or tracer. Once this equilibrium has been established, any subsequent loss of radiostrontium can be accurately assessed through yield determination of carrier or tracer.
It is important, therefore, that all. steps leading to equilibration be performed .with extra care to avoid losses in these steps.
If a significant amount (several milligrams) of stable strontium is present in the sample, as is the case for sea water, some shellfish, and certain soils, the gravimetric yield determination will be in error if a correction is not made for the additional strontium.
Most other bioenvironmental media of concern have small concentrations of stable- strontium which are not expected to cause a problem and in general can be ignored.
Calcium, however, is present in significant amounts in most bioenvironmental media, and its' separation from small amounts of added strontium carrier can be tedious;
if this separation is not performed properly, however, the gravimetric yield will be biased in proportion to the amount of calcium remaining in the strontium fraction.
Since Sr-90 and Sr-89 concentrations in normal bioenvironmental media are expected to be low,
-appropriate low-level techniques should be used in making these measurements. Some of the better high precision radiochemical procedures for low-level Sr-90
assay depend on counting the Y-90 daughter activity,.
rather than Sr-90 itself or the combined Sr-90 - Y-90
activities. There are several advantages of the former technique. The 64-hour half-life of Y-90 makes it convenient for following radioactive decay as.a check on isotopic purity, and the single-component activity minimizes ambiguities in its measurement. Strontium-89 determinations are made by difference, i.e., by taking the difference between total strontium counts and the counts due to Sr-90 alone.
The Sr-89 and Sr-90 radioanalytical procedures described in HASL 300 ' and SWRHL-l1
2 have been in routine use for many years at their respective laboratories and have been. shown to provide consistently good results in terms of both sensitivity and accuracy.
C. REGULATORY POSITION
The analytical procedures for Sr-89 and Sr-90
described in HASL 300 ' and in SWRHL-1 1 2 are acceptable to the Regulatory staff as bases for meeting the analytical performance standards needed to assess accurately the Sr-89 and Sr-90 levels in biological and environmental media.