Last Measurement Q.A. Report 4/18/17 8:02:09 AM Page 1

          • G E N I E Q U A L I T Y A S S U R A N C E *****

Last Results Report 4/18/17 8:02:09 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: UNC Daily QC 041 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 4/18/17 7:50:01 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 352.1 seconds Parameter Description Value Deviation/Flags

[Mean +/- Std. Dev.] < LU : SD : UD : BS >

Peak FWHM (121.8 keV) 9.6095E-001 1.6886E-002

[SD: 9.6073E-001+/- 0.013] < : : : >

Peak FWHM (1407.92 keV) 2.2666E+000 -7.8801E-001

[SD: 2.3620E+000+/- 0.121] < : : : >

Peak FWHM (964.13 keV) 1.8077E+000 -1.3683E+000

[SD: 1.9132E+000+/- 0.077] < : : : >

Peak energy (964.13 keV) 9.6397E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (121.8 keV) 1.2192E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (1407.92 keV 1.4082E+003 0.00

[0.00+/-0.00] < : : : >

Corrected activity (121. 5.4687E-001 -4.8047E-001

[SD: 5.5013E-001+/- 0.007] < : : : >

Corrected activity (964. 4.8419E-001 -2.2132E-002

[SD: 4.8437E-001+/- 0.008] < : : : >

Corrected activity (1407 4.8533E-001 2.1274E-003

[SD: 4.8532E-001+/- 0.008] < : : : >

Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)

SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)

UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)

BS = Measurement Bias Test (In = Investigate, Ac = Action)