Last Measurement Q.A. Report 4/26/17 7:42:37 AM Page 1
- G E N I E Q U A L I T Y A S S U R A N C E *****
Last Results Report 4/26/17 7:42:37 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: UNC Daily QC 042 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 4/26/17 7:34:07 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 351.9 seconds Parameter Description Value Deviation/Flags
[Mean +/- Std. Dev.] < LU : SD : UD : BS >
Peak FWHM (121.8 keV) 9.6509E-001 2.6548E-001
[SD: 9.6171E-001+/- 0.013] < : : : >
Peak FWHM (1407.92 keV) 2.5826E+000 1.6235E+000
[SD: 2.3709E+000+/- 0.130] < : : : >
Peak FWHM (964.13 keV) 1.9727E+000 8.1349E-001
[SD: 1.9084E+000+/- 0.079] < : : : >
Peak energy (964.13 keV) 9.6408E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (121.8 keV) 1.2195E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (1407.92 keV 1.4084E+003 0.00
[0.00+/-0.00] < : : : >
Corrected activity (121. 5.4708E-001 -4.2025E-001
[SD: 5.4975E-001+/- 0.006] < : : : >
Corrected activity (964. 4.9245E-001 1.0977E+000
[SD: 4.8434E-001+/- 0.007] < : : : >
Corrected activity (1407 4.8234E-001 -4.0908E-001
[SD: 4.8527E-001+/- 0.007] < : : : >
Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)
SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)
UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)
BS = Measurement Bias Test (In = Investigate, Ac = Action)