Last Measurement Q.A. Report 5/16/17 8:43:18 AM Page 1
- G E N I E Q U A L I T Y A S S U R A N C E *****
Last Results Report 5/16/17 8:43:18 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: Daily QC 0516201 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 12:00:00 AM Measurement Date: 5/16/17 8:32:08 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 351.7 seconds Parameter Description Value Deviation/Flags
[Mean +/- Std. Dev.] < LU : SD : UD : BS >
Peak FWHM (121.8 keV) 9.6206E-001 2.2967E-002
[SD: 9.6180E-001+/- 0.012] < : : : >
Peak FWHM (1407.92 keV) 2.4693E+000 5.6382E-001
[SD: 2.3898E+000+/- 0.141] < : : : >
Peak FWHM (964.13 keV) 2.0284E+000 1.3325E+000
[SD: 1.9180E+000+/- 0.083] < : : : >
Peak energy (964.13 keV) 9.6421E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (121.8 keV) 1.2197E+002 0.00
[0.00+/-0.00] < : : : >
Peak energy (1407.92 keV 1.4086E+003 0.00
[0.00+/-0.00] < : : : >
Corrected activity (121. 5.4275E-001 -3.5903E-001
[SD: 5.4710E-001+/- 0.012] < : : : >
Corrected activity (964. 4.8194E-001 -3.4437E-001
[SD: 4.8431E-001+/- 0.007] < : : : >
Corrected activity (1407 4.8838E-001 4.5885E-001
[SD: 4.8513E-001+/- 0.007] < : : : >
Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)
SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)
UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)