Last Measurement Q.A. Report 6/20/17 10:46:29 AM Page 1

          • G E N I E Q U A L I T Y A S S U R A N C E *****

Last Results Report 6/20/17 10:46:29 AM QA File: C:\\GENIE2K\\CAMFILES\\UNC 2017 856 Sample ID: Daily QC 062020 Sample Quantity: 1.0000E+000 Sample Date: 6/1/11 2:00:00 PM Measurement Date: 6/20/17 10:28:28 AM Elapsed Live Time: 300.0 seconds Elapsed Real Time: 349.4 seconds Parameter Description Value Deviation/Flags

[Mean +/- Std. Dev.] < LU : SD : UD : BS >

Peak FWHM (121.8 keV) 9.4743E-001 -1.2345E+000

[SD: 9.6388E-001+/- 0.013] < : : : >

Peak FWHM (1407.92 keV) 2.3179E+000 -5.0014E-001

[SD: 2.3910E+000+/- 0.146] < : : : >

Peak FWHM (964.13 keV) 1.8550E+000 -7.6002E-001

[SD: 1.9212E+000+/- 0.087] < : : : >

Peak energy (964.13 keV) 9.6409E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (121.8 keV) 1.2193E+002 0.00

[0.00+/-0.00] < : : : >

Peak energy (1407.92 keV 1.4084E+003 0.00

[0.00+/-0.00] < : : : >

Corrected activity (121. 5.4285E-001 -2.9360E-001

[SD: 5.4630E-001+/- 0.012] < : : : >

Corrected activity (964. 4.8846E-001 7.0621E-001

[SD: 4.8341E-001+/- 0.007] < : : : >

Corrected activity (1407 4.8488E-001 -2.7623E-002

[SD: 4.8507E-001+/- 0.007] < : : : >

Flags Key: LU = Lower/Upper Bounds Test (Ab = Above, Be = Below)

SD = Sample Driven N-Sigma Test (In = Investigate, Ac = Action)

UD = User Driven N-Sigma Test (In = Investigate, Ac = Action)

BS = Measurement Bias Test (In = Investigate, Ac = Action)