Regulatory Guide 8.14: Difference between revisions

From kanterella
Jump to navigation Jump to search
(Created page by program invented by StriderTol)
(Created page by program invented by StriderTol)
Line 14: Line 14:
| page count = 3
| page count = 3
}}
}}
{{#Wiki_filter:U.S. NUCLEAR REGULATORY COMMISSIONRevision IAugust 1977REGULATORY GUOIDEOFFICE OF STANDARDS DEVELOPMENTREGULATORY GUIDE 8.14PERSONNEL NEUTRON DOSIMETERS
{{#Wiki_filter:U.S. NUCLEAR REGULATORY  
COMMISSION
Revision I August 1977 REGULATORY
GUOIDE OFFICE OF STANDARDS  
DEVELOPMENT
REGULATORY
GUIDE 8.14 PERSONNEL
NEUTRON DOSIMETERS


==A. INTRODUCTION==
==A. INTRODUCTION==
Section 20.202, "Personnel Monitoring," of 10CFR Part 20, "Standards for Protection AgainstRadiation," requires that licensees supply personnelmonitoring equipment and require its use by specifiedindividuals. This guide provides guidance acceptableto the NRC staff on the use of personnel neutrondosimeters where exposure to neutrons occurs.
Section 20.202, "Personnel Monitoring," of 10 CFR Part 20, "Standards for Protection Against Radiation," requires that licensees supply personnel monitoring equipment and require its use by specified individuals.
 
This guide provides guidance acceptable to the NRC staff on the use of personnel neutron dosimeters where exposure to neutrons occurs.


==B. DISCUSSION==
==B. DISCUSSION==
The American National Standards Institute hasapproved a standard entitled "Personnel NeutronDosimeters. (Neutron Energies Less than 20 MeV)"and designated ANSI N319-1976.' This standardgives performance criteria, use factors, and dosimetrysystem calibration criteria for neutron dosimetrysystems. Accuracy criteria are not included amongthe performance criteria, however.This guide supplements the standard by adding anaccuracy requirement. The guide also provides sub-stitutes for certain of the standard's performance re-quirements because data from a test performed by theBattelle Pacific Northwest Laboratory showed thatthe requirements of the standard could not be met byNTA film or thermoluminescence dosimeter systems.A copy of the Battelle data is available by writing tothe Occupational Health Standards Branch, Office ofStandards Development, U.S. Nuclear RegulatoryCommission, Washington, D.C. 20555.One problem in neutron dosimetry is that neutrondosimeters are less sensitive than gamma dosimeters.Section 20.202 of the NRC regulations requires per-sonnel monitoring if a worker is likely to receive a'Copies may be obtained from the American National StandardsInstitute, Inc., 1430 Broadway, New York, New York 10018.*Lines indicate substantive changes from previous issue.whole body dose of more than about 300 millirems ina quarter year. Many neutron dosimeters have dif-ficulty measuring neutron doses of this magnitude.The problem is even more severe in a mixed radiationfield where neutrons contribute only a portion of the300 millirems. Thus the regulations may require per-sonnel monitoring when the neutron dose is less than300 millirems in a quarter, but the standard only re-quires the dosimeters to detect 300 millirems perquarter (paragraph 4.1 of the standard). Thisproblem is discussed in regulatory position C.I.
The American National Standards Institute has approved a standard entitled "Personnel Neutron Dosimeters. (Neutron Energies Less than 20 MeV)" and designated ANSI N319-1976.'  
This standard gives performance criteria, use factors, and dosimetry system calibration criteria for neutron dosimetry systems. Accuracy criteria are not included among the performance criteria, however.This guide supplements the standard by adding an accuracy requirement.
 
The guide also provides sub-stitutes for certain of the standard's performance re-quirements because data from a test performed by the Battelle Pacific Northwest Laboratory showed that the requirements of the standard could not be met by NTA film or thermoluminescence dosimeter systems.A copy of the Battelle data is available by writing to the Occupational Health Standards Branch, Office of Standards Development, U.S. Nuclear Regulatory Commission, Washington, D.C. 20555.One problem in neutron dosimetry is that neutron dosimeters are less sensitive than gamma dosimeters.
 
Section 20.202 of the NRC regulations requires per-sonnel monitoring if a worker is likely to receive a'Copies may be obtained from the American National Standards Institute, Inc., 1430 Broadway, New York, New York 10018.*Lines indicate substantive changes from previous issue.whole body dose of more than about 300 millirems in a quarter year. Many neutron dosimeters have dif-ficulty measuring neutron doses of this magnitude.
 
The problem is even more severe in a mixed radiation field where neutrons contribute only a portion of the 300 millirems.
 
Thus the regulations may require per-sonnel monitoring when the neutron dose is less than 300 millirems in a quarter, but the standard only re-quires the dosimeters to detect 300 millirems per quarter (paragraph  
4.1 of the standard).  
This problem is discussed in regulatory position C.I.C. REGULATORY
POSITION ANSI N319-1976, "Personnel Neutron Dosi-meters (Neutron Energies Less than 20 MeV),"'provides guidance acceptable to the NRC staff on the use of personnel neutron dosimeters, as sup-plemented and modified below.1. When Neutron Dosimeters Should Be Worn Neutron dosimeters should be worn whenever the neutron dose equivalent is likely to exceed 300 mrem in a quarter (the minimum sensitivity required of a neutron dosimeter in paragraph
4.1 of the standard).
If personnel monitoring is required by §20.202 because of the total radiation exposure but the neutron dose equivalent is not likely to exceed 300 mrem in a quarter, the following alternatives are ac-ceptable with regard to personnel neutron dosimetry:
a. Using higher sensitivity dosimeters.
 
The licensee may use a dosimeter with a sensitivity greater than that required in the standard.
 
Albedo neutron dosimeters generally are believed to be more sensitive than required by the standard.
 
Film may be accep-table if fading due to humidity can be controlled, if the neutron spectrum has a small proportion of neutrons below the neutron energy threshold of the film (about 0.7 MeV), and if the tracks are counted on a large area of film.USNRC-REGULATORY
GUIDES Comments shou!d be sent to the Secretary of the Commission, US. Nuclear Regs.latory Commission, Washington, D.C. 20555, Attention:
Docketing and Service Regulatory Guides are issued to describe and make available to the public methods rranch.acceptable to the N RC staff of implementing specific Darts of the Commission's reeulatio-s, to delineate techniques used by the staff in evaluating specific ,"roblemns The guides are issued in the following ton broad divisions:
or Postut3ted accidents, or to provide guidance to applicants.
 
Regulatory Guides arc not subslitutes for regulations, and compliance with them is not required.
 
1. Power Reactors 6. Products Methods and soiullons diflerent from those set out in the nuides will be accept. 2. Research and Test Reactors 7. Transportation able if they provide a basis for the findings requisite to the isruance or continuance
3. Fuels and Materials Facilities
8. Occupational Health of a permit or license by the Commission
4. Environmental end Siting 9. Antitrust Review 5. Materials and Plant Protection
10. General Comments and suggestions for improvements in these guides ar.
 
at all Requests for single copies of issued guides lwhich may be reoroduced)
or for place-times; and guides will be revised, as appropriate, to accommodate cominents and rlent on an automatic dristribution fist tor single copies o! future guides in Specific to reflect new information or experience.
 
This guide was revisad as a result of divisions should be made in writing to the US. Nuclear Regulatory Commission, substantive comment; receved from the Public and additional staff revyew. Washington.
 
D.C. 20555, Attention:
Director, Division of Document Control.
 
b. Calculated neutron dose equivalent to supplement neutron dosimeter.
 
A licensee may use a personnel neutron dosimeter but may substitute a calculated neutron dose equivalent for the measured dose equivalent if the measured dose equivalent cannot be reliably determined because of the lack of sensitivity of the dosimeter.
 
Calculated dose equivalents may be based on measured neutron/gamma ratios or on neutron dose equivalents measured with portable or fixed monitoring instruments and known personnel occupancy times. More information on determining neutron dose equivalents by neutron/gamma ratios is given in Regulatory Guide 8.4, "Direct-Reading and indirect-Reading Pocket Dosimeters." c. Calculated neutron dose equivalent in place of neutron dosimeter.
 
If the individual is not likely to receive a neutron dose equivalent in excess of 100 mrem in a quarter but wouid still have to have some sort of monitoring under §20.202 (e.g., gamma monitoring), a personnel neutron dosimeter may be omitted. The neutron dose equivalent should then be estimated by the methods in regulatory position C.L.b above. This procedure is discussed in more detail in regulatory position C.3 of Regulatory Guide 8.4.d. Neutron dose equivalent much smaller than gam-ma dose equivalent.
 
If the neutron whole body dose equivalent is not likely to exceed 10% of the gamma plus x-ray dose equivalent, neutron dosimeters may be omitted and the neutron dose equivalent may be assumed equal to zero.e. Negligible neutron dose equivalent.
 
If the neutron dose equivalent is not likely to exceed 30 mrem per quarter, or 10 mrem per quarter for in-dividuals under 18 years of age, neutron dosimeters may be omitted and the neutron dose equivalent as-sumed equal to zero. The determination that an in-dividual is not likely to receive a neutron dose equivalent of 30 mrem per quarter should not be based on previous NTA film badge readings since NTA film is not suffficiently sensitive to detect this dose equivalent rate.2. Performance Requirements Section 4 of the standard contains -several perfor-mance requirements on the dosimetry system.Instead of paragraph
4.1 of the standard on lower limit of detection, the following should be used: "The lower limit of detection of the dosimetry system shall not exceed 300 mrem per quarter. The quarterly lower limit of detection (LLD) in mrems at the 95% confidence level is defined as LLD = 4.66/s S where N is the number of dosimetry exchange periods in a quarter and S is the standard deviation of the normal background in mrems for a single reading." (This definition of LLD was chosen to be consistent with the NRC position previously stated in Tables 1 and 3 of Regulatory Guide 4.8, "Environmental Technical Specifications for Nuclear Power Plants." The basis for the definition is given in USERDA report HASL-300, p. D-08-01.)Instead of the 10% limit on the standard deviation in paragraph
4.4 of the standard on precision, a limit of 30% should be used.The following accuracy requirement should be ad-ded: "When exposed to an unmoderated californium-
252 source; the average accuracy of a set of 10 dosimeters exposed in the range from 100 mrems to 3 rems should be +/-50%." The tests necessary to verify that the dosimetry system meets the requirements in the standard may be performed by the licensee or by someone selected by him. The tests may be performed on a system just once to demonstrate that it can meet the require-ments of Section 4 of the standard.In the case of film, each different type of packaging should be considered a different system to be tested separately.
 
Tests should also be repeated any time changes are made in the dosimetry system (i.e., processing the dosimeters differently).
A licensee fol-lowing the recommendations of this guide should maintain records to show that his neutron dosimeters have been tested and meet the perforrriance require-ments of Section 4 of the standard.3. Meeting the Fading Requirements of the Standard.To meet the fading requirements in paragraphs
5.3 and 5.4 of the standard, the licensee should irradiate four dosimeters with a neutron dose equivalent of at least 0.5 rem. The dosimeters should then be stored for an entire dosimeter exchange period at a humidity similar to that expected in normal use. The dosimeters should then be read to verify that, on the average, they meet the fading criterion (paragraph
5.3 of the standard).
If humidity is a significant cause of fading, the test should be performed during a period of high humidity.4. Dosimetry System Calibration Section 6 of the standard gives suggestions on dosimetry system calibration.
 
Calibration of dosimeter response may. be performed by the ex-posure of the dosimeter and a reference monitoring instrument in the actual locations where significant neutron exposure occurs. The spherical (Hankins)"rem-meter," the Anderson-Braun rem counter, or other similar instruments may be -used as reference monitoring instruments.
 
These monitoring instruýments should be calibrated by sources whose calibra-tion is traceable to the National Bureau of Standards.
 
I 8. 14-2 Calibrations of dosimeter response may also be performed by the dosimeter processor.
 
The processor should use a calibration factor applicable to the dose spectrum delivered to the individuals wearing the dosimeters.
 
The calibration may be performed by the licensee or by someone selected by him, and records describing the dosimeter calibration should be main-tained by the licensee.


==C. REGULATORY POSITION==
The dosimeter calibration should be repeated any time the licensee has reason to believe the neutron spectrum has changed and the previous calibration might not be valid.5. Maintaining Body Contact Albedo neutron systems usually require close body contact at all times during usage. Sizable errors can occur if close body contact is not maintained.
ANSI N319-1976, "Personnel Neutron Dosi-meters (Neutron Energies Less than 20 MeV),"'provides guidance acceptable to the NRC staff on theuse of personnel neutron dosimeters, as sup-plemented and modified below.1. When Neutron Dosimeters Should Be WornNeutron dosimeters should be worn whenever theneutron dose equivalent is likely to exceed 300 mremin a quarter (the minimum sensitivity required of aneutron dosimeter in paragraph 4.1 of the standard).If personnel monitoring is required by §20.202because of the total radiation exposure but theneutron dose equivalent is not likely to exceed 300mrem in a quarter, the following alternatives are ac-ceptable with regard to personnel neutron dosimetry:a. Using higher sensitivity dosimeters. The licenseemay use a dosimeter with a sensitivity greater thanthat required in the standard. Albedo neutrondosimeters generally are believed to be more sensitivethan required by the standard. Film may be accep-table if fading due to humidity can be controlled, ifthe neutron spectrum has a small proportion ofneutrons below the neutron energy threshold of thefilm (about 0.7 MeV), and if the tracks are countedon a large area of film.USNRC-REGULATORY GUIDES Comments shou!d be sent to the Secretary of the Commission, US. Nuclear Regs.latory Commission, Washington, D.C. 20555, Attention: Docketing and ServiceRegulatory Guides are issued to describe and make available to the public methods rranch.acceptable to the N RC staff of implementing specific Darts of the Commission'sreeulatio-s, to delineate techniques used by the staff in evaluating specific ,"roblemns The guides are issued in the following ton broad divisions:or Postut3ted accidents, or to provide guidance to applicants. Regulatory Guidesarc not subslitutes for regulations, and compliance with them is not required. 1. Power Reactors 6. ProductsMethods and soiullons diflerent from those set out in the nuides will be accept. 2. Research and Test Reactors 7. Transportationable if they provide a basis for the findings requisite to the isruance or continuance 3. Fuels and Materials Facilities 8. Occupational Healthof a permit or license by the Commission 4. Environmental end Siting 9. Antitrust Review5. Materials and Plant Protection 10. GeneralComments and suggestions for improvements in these guides ar. at all Requests for single copies of issued guides lwhich may be reoroduced) or for place-times; and guides will be revised, as appropriate, to accommodate cominents and rlent on an automatic dristribution fist tor single copies o! future guides in Specificto reflect new information or experience. This guide was revisad as a result of divisions should be made in writing to the US. Nuclear Regulatory Commission,substantive comment; receved from the Public and additional staff revyew. Washington. D.C. 20555, Attention: Director, Division of Document Control.


b. Calculated neutron dose equivalent to supplementneutron dosimeter. A licensee may use a personnelneutron dosimeter but may substitute a calculatedneutron dose equivalent for the measured doseequivalent if the measured dose equivalent cannot bereliably determined because of the lack of sensitivityof the dosimeter. Calculated dose equivalents may bebased on measured neutron/gamma ratios or onneutron dose equivalents measured with portable orfixed monitoring instruments and known personneloccupancy times. More information on determiningneutron dose equivalents by neutron/gamma ratios isgiven in Regulatory Guide 8.4, "Direct-Reading andindirect-Reading Pocket Dosimeters."c. Calculated neutron dose equivalent in place ofneutron dosimeter. If the individual is not likely toreceive a neutron dose equivalent in excess of 100mrem in a quarter but wouid still have to have somesort of monitoring under §20.202 (e.g., gammamonitoring), a personnel neutron dosimeter may beomitted. The neutron dose equivalent should then beestimated by the methods in regulatory positionC.L.b above. This procedure is discussed in moredetail in regulatory position C.3 of Regulatory Guide8.4.d. Neutron dose equivalent much smaller than gam-ma dose equivalent. If the neutron whole body doseequivalent is not likely to exceed 10% of the gammaplus x-ray dose equivalent, neutron dosimeters maybe omitted and the neutron dose equivalent may beassumed equal to zero.e. Negligible neutron dose equivalent. If theneutron dose equivalent is not likely to exceed 30mrem per quarter, or 10 mrem per quarter for in-dividuals under 18 years of age, neutron dosimetersmay be omitted and the neutron dose equivalent as-sumed equal to zero. The determination that an in-dividual is not likely to receive a neutron doseequivalent of 30 mrem per quarter should not bebased on previous NTA film badge readings sinceNTA film is not suffficiently sensitive to detect thisdose equivalent rate.2. Performance RequirementsSection 4 of the standard contains -several perfor-mance requirements on the dosimetry system.Instead of paragraph 4.1 of the standard on lowerlimit of detection, the following should be used:"The lower limit of detection of the dosimetrysystem shall not exceed 300 mrem per quarter. Thequarterly lower limit of detection (LLD) in mrems atthe 95% confidence level is defined asLLD = 4.66/s Swhere N is the number of dosimetry exchange periodsin a quarter and S is the standard deviation of thenormal background in mrems for a single reading."(This definition of LLD was chosen to be consistentwith the NRC position previously stated in Tables 1and 3 of Regulatory Guide 4.8, "EnvironmentalTechnical Specifications for Nuclear Power Plants."The basis for the definition is given in USERDAreport HASL-300, p. D-08-01.)Instead of the 10% limit on the standard deviationin paragraph 4.4 of the standard on precision, a limitof 30% should be used.The following accuracy requirement should be ad-ded: "When exposed to an unmoderated californium-252 source; the average accuracy of a set of 10dosimeters exposed in the range from 100 mrems to 3rems should be +/-50%."The tests necessary to verify that the dosimetrysystem meets the requirements in the standard maybe performed by the licensee or by someone selectedby him. The tests may be performed on a system justonce to demonstrate that it can meet the require-ments of Section 4 of the standard.In the case of film, each different type of packagingshould be considered a different system to be testedseparately. Tests should also be repeated any timechanges are made in the dosimetry system (i.e.,processing the dosimeters differently). A licensee fol-lowing the recommendations of this guide shouldmaintain records to show that his neutron dosimetershave been tested and meet the perforrriance require-ments of Section 4 of the standard.3. Meeting the Fading Requirements of the Standard.To meet the fading requirements in paragraphs 5.3and 5.4 of the standard, the licensee should irradiatefour dosimeters with a neutron dose equivalent of atleast 0.5 rem. The dosimeters should then be storedfor an entire dosimeter exchange period at a humiditysimilar to that expected in normal use. Thedosimeters should then be read to verify that, on theaverage, they meet the fading criterion (paragraph 5.3of the standard). If humidity is a significant cause offading, the test should be performed during a periodof high humidity.4. Dosimetry System CalibrationSection 6 of the standard gives suggestions ondosimetry system calibration. Calibration ofdosimeter response may. be performed by the ex-posure of the dosimeter and a reference monitoringinstrument in the actual locations where significantneutron exposure occurs. The spherical (Hankins)"rem-meter," the Anderson-Braun rem counter, orother similar instruments may be -used as referencemonitoring instruments. These monitoring instruýments should be calibrated by sources whose calibra-tion is traceable to the National Bureau of Standards.I8. 14-2 Calibrations of dosimeter response may also beperformed by the dosimeter processor. The processorshould use a calibration factor applicable to the dosespectrum delivered to the individuals wearing thedosimeters. The calibration may be performed by thelicensee or by someone selected by him, and recordsdescribing the dosimeter calibration should be main-tained by the licensee. The dosimeter calibrationshould be repeated any time the licensee has reason tobelieve the neutron spectrum has changed and theprevious calibration might not be valid.5. Maintaining Body ContactAlbedo neutron systems usually require close bodycontact at all times during usage. Sizable errors canoccur if close body contact is not maintained. Albedoneutron dosimeters should have a means to maintainthis necessary close contact with the body.
Albedo neutron dosimeters should have a means to maintain this necessary close contact with the body.


==D. IMPLEMENTATION==
==D. IMPLEMENTATION==
The purpose of this section is to provide informa-tion to applicants regarding the NRC staff's plans forusing this regulatory guide.Except in those cases in which the applicantproposes an acceptable alternative method for com-plying with §20.202 of the Commission's regulations,the method herein will be used in the evaluation ofsubmittals in connection with license applicationsdocketed after November 1, 1977.If an applicant wishes to use this regulatory guidein developing submittals for applications docketed onor before November 1, 1977, the pertinent portions ofthe application will be evaluated on the basis of thisguide.8.14-3  
The purpose of this section is to provide informa-tion to applicants regarding the NRC staff's plans for using this regulatory guide.Except in those cases in which the applicant proposes an acceptable alternative method for com-plying with §20.202 of the Commission's regulations, the method herein will be used in the evaluation of submittals in connection with license applications docketed after November 1, 1977.If an applicant wishes to use this regulatory guide in developing submittals for applications docketed on or before November 1, 1977, the pertinent portions of the application will be evaluated on the basis of this guide.8.14-3}}
}}


{{RG-Nav}}
{{RG-Nav}}

Revision as of 16:20, 18 July 2018

Revison 1, Personnel Neutron Dosimeters.
ML13064A088
Person / Time
Issue date: 08/31/1977
From:
NRC/OSD, Office of Nuclear Regulatory Research
To:
References
RG-8.014, Rev 1
Download: ML13064A088 (3)


U.S. NUCLEAR REGULATORY

COMMISSION

Revision I August 1977 REGULATORY

GUOIDE OFFICE OF STANDARDS

DEVELOPMENT

REGULATORY

GUIDE 8.14 PERSONNEL

NEUTRON DOSIMETERS

A. INTRODUCTION

Section 20.202, "Personnel Monitoring," of 10 CFR Part 20, "Standards for Protection Against Radiation," requires that licensees supply personnel monitoring equipment and require its use by specified individuals.

This guide provides guidance acceptable to the NRC staff on the use of personnel neutron dosimeters where exposure to neutrons occurs.

B. DISCUSSION

The American National Standards Institute has approved a standard entitled "Personnel Neutron Dosimeters. (Neutron Energies Less than 20 MeV)" and designated ANSI N319-1976.'

This standard gives performance criteria, use factors, and dosimetry system calibration criteria for neutron dosimetry systems. Accuracy criteria are not included among the performance criteria, however.This guide supplements the standard by adding an accuracy requirement.

The guide also provides sub-stitutes for certain of the standard's performance re-quirements because data from a test performed by the Battelle Pacific Northwest Laboratory showed that the requirements of the standard could not be met by NTA film or thermoluminescence dosimeter systems.A copy of the Battelle data is available by writing to the Occupational Health Standards Branch, Office of Standards Development, U.S. Nuclear Regulatory Commission, Washington, D.C. 20555.One problem in neutron dosimetry is that neutron dosimeters are less sensitive than gamma dosimeters.

Section 20.202 of the NRC regulations requires per-sonnel monitoring if a worker is likely to receive a'Copies may be obtained from the American National Standards Institute, Inc., 1430 Broadway, New York, New York 10018.*Lines indicate substantive changes from previous issue.whole body dose of more than about 300 millirems in a quarter year. Many neutron dosimeters have dif-ficulty measuring neutron doses of this magnitude.

The problem is even more severe in a mixed radiation field where neutrons contribute only a portion of the 300 millirems.

Thus the regulations may require per-sonnel monitoring when the neutron dose is less than 300 millirems in a quarter, but the standard only re-quires the dosimeters to detect 300 millirems per quarter (paragraph

4.1 of the standard).

This problem is discussed in regulatory position C.I.C. REGULATORY

POSITION ANSI N319-1976, "Personnel Neutron Dosi-meters (Neutron Energies Less than 20 MeV),"'provides guidance acceptable to the NRC staff on the use of personnel neutron dosimeters, as sup-plemented and modified below.1. When Neutron Dosimeters Should Be Worn Neutron dosimeters should be worn whenever the neutron dose equivalent is likely to exceed 300 mrem in a quarter (the minimum sensitivity required of a neutron dosimeter in paragraph

4.1 of the standard).

If personnel monitoring is required by §20.202 because of the total radiation exposure but the neutron dose equivalent is not likely to exceed 300 mrem in a quarter, the following alternatives are ac-ceptable with regard to personnel neutron dosimetry:

a. Using higher sensitivity dosimeters.

The licensee may use a dosimeter with a sensitivity greater than that required in the standard.

Albedo neutron dosimeters generally are believed to be more sensitive than required by the standard.

Film may be accep-table if fading due to humidity can be controlled, if the neutron spectrum has a small proportion of neutrons below the neutron energy threshold of the film (about 0.7 MeV), and if the tracks are counted on a large area of film.USNRC-REGULATORY

GUIDES Comments shou!d be sent to the Secretary of the Commission, US. Nuclear Regs.latory Commission, Washington, D.C. 20555, Attention:

Docketing and Service Regulatory Guides are issued to describe and make available to the public methods rranch.acceptable to the N RC staff of implementing specific Darts of the Commission's reeulatio-s, to delineate techniques used by the staff in evaluating specific ,"roblemns The guides are issued in the following ton broad divisions:

or Postut3ted accidents, or to provide guidance to applicants.

Regulatory Guides arc not subslitutes for regulations, and compliance with them is not required.

1. Power Reactors 6. Products Methods and soiullons diflerent from those set out in the nuides will be accept. 2. Research and Test Reactors 7. Transportation able if they provide a basis for the findings requisite to the isruance or continuance

3. Fuels and Materials Facilities

8. Occupational Health of a permit or license by the Commission

4. Environmental end Siting 9. Antitrust Review 5. Materials and Plant Protection

10. General Comments and suggestions for improvements in these guides ar.

at all Requests for single copies of issued guides lwhich may be reoroduced)

or for place-times; and guides will be revised, as appropriate, to accommodate cominents and rlent on an automatic dristribution fist tor single copies o! future guides in Specific to reflect new information or experience.

This guide was revisad as a result of divisions should be made in writing to the US. Nuclear Regulatory Commission, substantive comment; receved from the Public and additional staff revyew. Washington.

D.C. 20555, Attention:

Director, Division of Document Control.

b. Calculated neutron dose equivalent to supplement neutron dosimeter.

A licensee may use a personnel neutron dosimeter but may substitute a calculated neutron dose equivalent for the measured dose equivalent if the measured dose equivalent cannot be reliably determined because of the lack of sensitivity of the dosimeter.

Calculated dose equivalents may be based on measured neutron/gamma ratios or on neutron dose equivalents measured with portable or fixed monitoring instruments and known personnel occupancy times. More information on determining neutron dose equivalents by neutron/gamma ratios is given in Regulatory Guide 8.4, "Direct-Reading and indirect-Reading Pocket Dosimeters." c. Calculated neutron dose equivalent in place of neutron dosimeter.

If the individual is not likely to receive a neutron dose equivalent in excess of 100 mrem in a quarter but wouid still have to have some sort of monitoring under §20.202 (e.g., gamma monitoring), a personnel neutron dosimeter may be omitted. The neutron dose equivalent should then be estimated by the methods in regulatory position C.L.b above. This procedure is discussed in more detail in regulatory position C.3 of Regulatory Guide 8.4.d. Neutron dose equivalent much smaller than gam-ma dose equivalent.

If the neutron whole body dose equivalent is not likely to exceed 10% of the gamma plus x-ray dose equivalent, neutron dosimeters may be omitted and the neutron dose equivalent may be assumed equal to zero.e. Negligible neutron dose equivalent.

If the neutron dose equivalent is not likely to exceed 30 mrem per quarter, or 10 mrem per quarter for in-dividuals under 18 years of age, neutron dosimeters may be omitted and the neutron dose equivalent as-sumed equal to zero. The determination that an in-dividual is not likely to receive a neutron dose equivalent of 30 mrem per quarter should not be based on previous NTA film badge readings since NTA film is not suffficiently sensitive to detect this dose equivalent rate.2. Performance Requirements Section 4 of the standard contains -several perfor-mance requirements on the dosimetry system.Instead of paragraph

4.1 of the standard on lower limit of detection, the following should be used: "The lower limit of detection of the dosimetry system shall not exceed 300 mrem per quarter. The quarterly lower limit of detection (LLD) in mrems at the 95% confidence level is defined as LLD = 4.66/s S where N is the number of dosimetry exchange periods in a quarter and S is the standard deviation of the normal background in mrems for a single reading." (This definition of LLD was chosen to be consistent with the NRC position previously stated in Tables 1 and 3 of Regulatory Guide 4.8, "Environmental Technical Specifications for Nuclear Power Plants." The basis for the definition is given in USERDA report HASL-300, p. D-08-01.)Instead of the 10% limit on the standard deviation in paragraph

4.4 of the standard on precision, a limit of 30% should be used.The following accuracy requirement should be ad-ded: "When exposed to an unmoderated californium-

252 source; the average accuracy of a set of 10 dosimeters exposed in the range from 100 mrems to 3 rems should be +/-50%." The tests necessary to verify that the dosimetry system meets the requirements in the standard may be performed by the licensee or by someone selected by him. The tests may be performed on a system just once to demonstrate that it can meet the require-ments of Section 4 of the standard.In the case of film, each different type of packaging should be considered a different system to be tested separately.

Tests should also be repeated any time changes are made in the dosimetry system (i.e., processing the dosimeters differently).

A licensee fol-lowing the recommendations of this guide should maintain records to show that his neutron dosimeters have been tested and meet the perforrriance require-ments of Section 4 of the standard.3. Meeting the Fading Requirements of the Standard.To meet the fading requirements in paragraphs

5.3 and 5.4 of the standard, the licensee should irradiate four dosimeters with a neutron dose equivalent of at least 0.5 rem. The dosimeters should then be stored for an entire dosimeter exchange period at a humidity similar to that expected in normal use. The dosimeters should then be read to verify that, on the average, they meet the fading criterion (paragraph

5.3 of the standard).

If humidity is a significant cause of fading, the test should be performed during a period of high humidity.4. Dosimetry System Calibration Section 6 of the standard gives suggestions on dosimetry system calibration.

Calibration of dosimeter response may. be performed by the ex-posure of the dosimeter and a reference monitoring instrument in the actual locations where significant neutron exposure occurs. The spherical (Hankins)"rem-meter," the Anderson-Braun rem counter, or other similar instruments may be -used as reference monitoring instruments.

These monitoring instruýments should be calibrated by sources whose calibra-tion is traceable to the National Bureau of Standards.

I 8. 14-2 Calibrations of dosimeter response may also be performed by the dosimeter processor.

The processor should use a calibration factor applicable to the dose spectrum delivered to the individuals wearing the dosimeters.

The calibration may be performed by the licensee or by someone selected by him, and records describing the dosimeter calibration should be main-tained by the licensee.

The dosimeter calibration should be repeated any time the licensee has reason to believe the neutron spectrum has changed and the previous calibration might not be valid.5. Maintaining Body Contact Albedo neutron systems usually require close body contact at all times during usage. Sizable errors can occur if close body contact is not maintained.

Albedo neutron dosimeters should have a means to maintain this necessary close contact with the body.

D. IMPLEMENTATION

The purpose of this section is to provide informa-tion to applicants regarding the NRC staff's plans for using this regulatory guide.Except in those cases in which the applicant proposes an acceptable alternative method for com-plying with §20.202 of the Commission's regulations, the method herein will be used in the evaluation of submittals in connection with license applications docketed after November 1, 1977.If an applicant wishes to use this regulatory guide in developing submittals for applications docketed on or before November 1, 1977, the pertinent portions of the application will be evaluated on the basis of this guide.8.14-3