Regulatory Guide 6.2: Difference between revisions
StriderTol (talk | contribs) (StriderTol Bot change) |
StriderTol (talk | contribs) (StriderTol Bot change) |
||
| (2 intermediate revisions by the same user not shown) | |||
| Line 1: | Line 1: | ||
{{Adams | {{Adams | ||
| number = | | number = ML003739394 | ||
| issue date = 07/31/1974 | | issue date = 07/31/1974 | ||
| title = Integrity and Test | | title = Rev 1,Integrity and Test Specifications for Selected Brachytherapy Sources | ||
| author name = | | author name = | ||
| author affiliation = | | author affiliation = NRC/RES | ||
| addressee name = | | addressee name = | ||
| addressee affiliation = | | addressee affiliation = | ||
| Line 10: | Line 10: | ||
| license number = | | license number = | ||
| contact person = | | contact person = | ||
| | | case reference number = RG-6.2 Rev 1 | ||
| document type = Regulatory Guide | | document type = Regulatory Guide | ||
| page count = 2 | | page count = 2 | ||
}} | }} | ||
{{#Wiki_filter: | {{#Wiki_filter:U.S. ATOMIC ENERGY COMMISSION | ||
REGULATORY | |||
DIRECTORATE OF REGULATORY STANDARDS | |||
Revision 1 July 1974 GUIDE | |||
REGULATORY GUIDE 6.2 INTEGRITY AND TEST SPECIFICATIONS FOR SELECTED BRACHYTHERAPY SOURCES | |||
==A. INTRODUCTION== | ==A. INTRODUCTION== | ||
Manulactuire of brachytherapy sources containing byproduct material is an activity requiring a license pursuant to section 30.3, "Activities Requiring License," | |||
Paragraph 32.74(a)(2)(iii) of 10 CFR Part 32, "Specific | l' 10) CFR Part 30, "Rules of General Applicability to Licensing of Byproduct Material." | ||
Licenses to Manufacture, Distribute, or Import Certain | Brachytherapy sources manufactured under such license must meet certain integrity requirements and pass certain tests. | ||
35.14, "Special Licenses for Certain Groups of Medical | |||
32 requires that the application also include details of | Paragraph 32.74(a)(2)(iii) of 10 CFR Part 32, "Specific Licenses to Manufacture, Distribute, or Import Certain Items Containing Byproduct Material," requires that an application for a specific license to manufacture and distribute brachytherapy sources and devices containing byproduct material to persons licensed under Section | ||
the quality control procedures which assure that pro | 35.14, "Special Licenses for Certain Groups of Medical Uses," of 10 CFR Part 35, "Human Uses of Byproduct Material," include a description of procedures for, and the results of, the prototype tests performed to demon strate that the source or device will maintain its integrity under stresses likely to be encountered in normal use and accidents. Paragraph 32.74(a)(2)(iv) of 10 CFR Part | ||
design and prototype tests. This guide describes integrity requirements and test specifications acceptable to the Regulatory staff for selected brachytherapy sources. | 32 requires that the application also include details of the quality control procedures which assure that pro duction sources and devices meet the standards of the design and prototype tests. This guide describes integrity requirements and test specifications' acceptable to the Regulatory staff for selected brachytherapy sources. | ||
==B. DISCUSSION== | ==B. DISCUSSION== | ||
type units tested pursuant to this standard." The quality control program further requires that, before transfer, | Subcommittee N44-2, Therapeutic Radiology, of the American National Standards Institute Committee on Equipment and Materials for Medical Radiation Appli cation, N44, has developed a standard presenting integ rity requirements and test specifications for selected | ||
* Line indicates substantive changes from previous issue. | |||
sealed medical sources used in interstitial, intracavitary, and topical therapy. This standard was approved by the American National Standards Institute (ANSI) | |||
on August 16, 1973, and designated ANSI N44.1-1973.1 The standard is limited to the traditional type sources with well-established medical uses and does not provide standards for sources used in afterloaded devices. The three source type categories covered by the standard are tubes or capsules and needles containing Ra-266, Co-60, | |||
or Cs-137, and beta applicators containing either Sr-90 | |||
or Ra-226. The standard establishes performance test specifications for temperature, impact, percussion, bend ing, tensile stress, and puncture. The temperature, impact, and percussion tests, which are the same tests for classification of a source as "special form" under the transportation regulations of the Department of Trans portation and the International Atomic Energy Agency are applicable to all source types. In addition, the bending test is applicable to interstitial needles, and the tensile stress and puncture tests are applicable to the beta applicators.. According to ANSI N44.1-1973, a source is deemed to pass these tests if it is "free of visual defects" and "demonstrates leakage of less than 50 | |||
nanocuries (nCi) (10 nanocuries of radon in 24 hours in the case of Ra-226 sources)." | |||
The standard also includes a quality control program | |||
"to insure that the production units will have the same integrity and meet the same requirements as the proto type units tested pursuant to this standard." The quality control program further requires that, before transfer, each source be tested and demonstrate leakage of less than 5 nanocuries (1 nanocurie of radon in 24 hours in the case of Ra-226 sources). | |||
1Copies may be obtained from the American National Standards Institute, Inc., 1430 Broadway, New York. New York 10018. | |||
USAEC REGULATORY GUIDES | USAEC REGULATORY GUIDES | ||
Copies of published guides may be obtained by request indicating the divisions desired to the U.S. Atomic Energy Commission, Washington, D.C. 20545, Regulatory Guides are issued to describe and make available to the public Attention: Director of Regulatory Standards. Comments and suggestions for methods acceptable to the AEC Regulatory staff of implementing specific parts of improvements in these guides are encouraged and should be sent to the Secretary the Commission's regulations, to delineate techniques | |||
-ed by the staff in of the Commission, U.S. Atomic Energy Commission, Washington, D.C. 20545, evaluating specific problems or postulated accidents, or to provide guidance to Attention: Chief, Public Proceedings Staff. | |||
applicants. Regulatory Guides are not substitutes for regulations and compliance with them is not required. Methods and solutions different from those set out in The guides are issued in the following ten broad divisions: | |||
the guides will be acceptable if they provide a basis for the findings requisite to the issuance or continuance of a permit or license by the Commission. | |||
===1. Power Reactors === | |||
6- Products | |||
2. Research and Test Reactors | |||
=== | ===7. Transportation === | ||
3. Fuels and Materials Facilities | |||
8. Occupational Health Published guides will be revised periodically, as appropriate, to accommodate | |||
4. Environmental and Siting | |||
9. Antitrust Review comments and to reflect new information or experience. | |||
5. Materials and Plant Protection | |||
1 | |||
===0. General=== | ===0. General=== | ||
==C. REGULATORY POSITION== | ==C. REGULATORY POSITION== | ||
materials | The requirements and recommended practices con tained in ANSI N44. 1-1973, "Integrity and Test Specifi cations for Selected Brachytherapy Sources," 1 con stitute a generally acceptable procedure foi demonstrating the integrity of those medical brachy therapy sources identified in the standard, subject to the following: | ||
1. Section 4.1 of ANSI N44.1-1973 states that non radioactive prototypes may be substituted for radio active units in cases where evidence of integrity can be established by leakage test methods appropriate t( | |||
nonradioactive sources. The integrity of the source will rw established if the source is tested by a method l:'e of accurately and reliably measuring leakage | |||
;,ies of 10- 7 atm cm3/sec or less for dry air at 25 0C and for a pressure differential of one atmospheric against a vacuum of 1/100 atmosphere or less, with negative results. | |||
2. Item 2 in Section 4.1 states that leakage of less than | |||
50 nanocuries from sources subjected to tests described in Section 4.2 denmonstrates retention of integrity after the test. Prototype sources subjected to tests described in Section 4.2 should retain their integrity and leak essentially no radioactive material. This will be demon strated for the purpose of these prototype tests, if leak tests performed both prior to and after each test described in Section 4.2 indicate removal of less than 5 nanocuries in each case. | |||
3. To the considerations described in Section 5.2, | |||
"Design Requirements," should be added the possible deleterious effects on source materials (both radioactive and nonradioactive) due to the inherent nature of the source and resulting from radiation damage, chemical and physical changes (e.g., those resulting in the evolu tion of gases or changes in density and volume of contained radioactive materials), and chemical reactions between source materials. | |||
4. Section 5.3, "Materials Specification and Control," | |||
states that the quality control program shall assure that | |||
"substantially the same .... | |||
materials .... | |||
are incorpo rated in all production units." (Emphasis supplied.) All production units should be produced with and contain materials with exactly the same characteristics as the materials used in the qualified prototype units. | |||
===5. Section === | |||
5.4 of ANSI N44.1-1973, | |||
"Acceptance Testing," states that the acceptability of the source shall be indicated by removal of less than 5 nanocuries of the radioisotope in one of the tests designed to demonstrate contamination of the outer capsule, and it recommends that a value not greater than 1/10 of the radioactivity limit above be chosen as a production control point and that no source which exceeds this value be transferred by a manfacturer to a user for use as a brachytherapy source. | |||
The acceptability of a source for transfer by a manufacturer to a user for use as a brachytherapy source should be indicated by removal of less than 0.5 nanocurie of the radioisotope in one of the tests designed to demonstrate contamination of the outer capsule and by presence of a hermetic seal according to a leakage test designed to demonstrate this situation. The vacuum leach tests, helium mass spectrometer tests, and Kr-85 leak tests 2 are acceptable for the purpose of demonstrating a hermetic seal. | |||
tests | 6. Manufacturers of brachytherapy sources should main tain results of prototype and acceptance tests. The test results should be expressed in terms of activity measured and include identification of the test method, instrumen tation, and calibration procedures used. | ||
Persons licensed by the AEC for byproduct, source, or | 7. The Atomic Energy Commission does riot license sources and devices containing Ra-226 and its daughters Persons licensed by the AEC for byproduct, source, or special nuclear material who also possess sources con taining radium and its daughters must limit the radiation exposure from both licensed material and other sources of radiation such as radium so that the limits in Part 20 | ||
are not exceeded (§20.101). Also, many of the States and all of the Agreement States do regulate the use of sources of radium and its daughters. For that reason, the parts of ANSI N44.1-1973 dealing specifically with testing of sealed radium brachytherapy sources are included in this guide, although the AEC has made no determination as to their adequacy. | |||
2 R.G. Neimeyer, | |||
"Leak Testing Encap.talted Radioactive Sources" ORNL-4529, July | |||
1972, available from National Technical Information Service, U.S. Department of Commerce, | |||
5285 Port Royal Road, Springfield, Va. 22151. | |||
6.2-2}} | 6.2-2}} | ||
{{RG-Nav}} | {{RG-Nav}} | ||
Latest revision as of 02:10, 17 January 2025
| ML003739394 | |
| Person / Time | |
|---|---|
| Issue date: | 07/31/1974 |
| From: | Office of Nuclear Regulatory Research |
| To: | |
| References | |
| RG-6.2 Rev 1 | |
| Download: ML003739394 (2) | |
U.S. ATOMIC ENERGY COMMISSION
REGULATORY
DIRECTORATE OF REGULATORY STANDARDS
Revision 1 July 1974 GUIDE
REGULATORY GUIDE 6.2 INTEGRITY AND TEST SPECIFICATIONS FOR SELECTED BRACHYTHERAPY SOURCES
A. INTRODUCTION
Manulactuire of brachytherapy sources containing byproduct material is an activity requiring a license pursuant to section 30.3, "Activities Requiring License,"
l' 10) CFR Part 30, "Rules of General Applicability to Licensing of Byproduct Material."
Brachytherapy sources manufactured under such license must meet certain integrity requirements and pass certain tests.
Paragraph 32.74(a)(2)(iii) of 10 CFR Part 32, "Specific Licenses to Manufacture, Distribute, or Import Certain Items Containing Byproduct Material," requires that an application for a specific license to manufacture and distribute brachytherapy sources and devices containing byproduct material to persons licensed under Section
35.14, "Special Licenses for Certain Groups of Medical Uses," of 10 CFR Part 35, "Human Uses of Byproduct Material," include a description of procedures for, and the results of, the prototype tests performed to demon strate that the source or device will maintain its integrity under stresses likely to be encountered in normal use and accidents. Paragraph 32.74(a)(2)(iv) of 10 CFR Part
32 requires that the application also include details of the quality control procedures which assure that pro duction sources and devices meet the standards of the design and prototype tests. This guide describes integrity requirements and test specifications' acceptable to the Regulatory staff for selected brachytherapy sources.
B. DISCUSSION
Subcommittee N44-2, Therapeutic Radiology, of the American National Standards Institute Committee on Equipment and Materials for Medical Radiation Appli cation, N44, has developed a standard presenting integ rity requirements and test specifications for selected
- Line indicates substantive changes from previous issue.
sealed medical sources used in interstitial, intracavitary, and topical therapy. This standard was approved by the American National Standards Institute (ANSI)
on August 16, 1973, and designated ANSI N44.1-1973.1 The standard is limited to the traditional type sources with well-established medical uses and does not provide standards for sources used in afterloaded devices. The three source type categories covered by the standard are tubes or capsules and needles containing Ra-266, Co-60,
or Cs-137, and beta applicators containing either Sr-90
or Ra-226. The standard establishes performance test specifications for temperature, impact, percussion, bend ing, tensile stress, and puncture. The temperature, impact, and percussion tests, which are the same tests for classification of a source as "special form" under the transportation regulations of the Department of Trans portation and the International Atomic Energy Agency are applicable to all source types. In addition, the bending test is applicable to interstitial needles, and the tensile stress and puncture tests are applicable to the beta applicators.. According to ANSI N44.1-1973, a source is deemed to pass these tests if it is "free of visual defects" and "demonstrates leakage of less than 50
nanocuries (nCi) (10 nanocuries of radon in 24 hours2.777778e-4 days <br />0.00667 hours <br />3.968254e-5 weeks <br />9.132e-6 months <br /> in the case of Ra-226 sources)."
The standard also includes a quality control program
"to insure that the production units will have the same integrity and meet the same requirements as the proto type units tested pursuant to this standard." The quality control program further requires that, before transfer, each source be tested and demonstrate leakage of less than 5 nanocuries (1 nanocurie of radon in 24 hours2.777778e-4 days <br />0.00667 hours <br />3.968254e-5 weeks <br />9.132e-6 months <br /> in the case of Ra-226 sources).
1Copies may be obtained from the American National Standards Institute, Inc., 1430 Broadway, New York. New York 10018.
USAEC REGULATORY GUIDES
Copies of published guides may be obtained by request indicating the divisions desired to the U.S. Atomic Energy Commission, Washington, D.C. 20545, Regulatory Guides are issued to describe and make available to the public Attention: Director of Regulatory Standards. Comments and suggestions for methods acceptable to the AEC Regulatory staff of implementing specific parts of improvements in these guides are encouraged and should be sent to the Secretary the Commission's regulations, to delineate techniques
-ed by the staff in of the Commission, U.S. Atomic Energy Commission, Washington, D.C. 20545, evaluating specific problems or postulated accidents, or to provide guidance to Attention: Chief, Public Proceedings Staff.
applicants. Regulatory Guides are not substitutes for regulations and compliance with them is not required. Methods and solutions different from those set out in The guides are issued in the following ten broad divisions:
the guides will be acceptable if they provide a basis for the findings requisite to the issuance or continuance of a permit or license by the Commission.
1. Power Reactors
6- Products
2. Research and Test Reactors
7. Transportation
3. Fuels and Materials Facilities
8. Occupational Health Published guides will be revised periodically, as appropriate, to accommodate
4. Environmental and Siting
9. Antitrust Review comments and to reflect new information or experience.
5. Materials and Plant Protection
1
0. General
C. REGULATORY POSITION
The requirements and recommended practices con tained in ANSI N44. 1-1973, "Integrity and Test Specifi cations for Selected Brachytherapy Sources," 1 con stitute a generally acceptable procedure foi demonstrating the integrity of those medical brachy therapy sources identified in the standard, subject to the following:
1. Section 4.1 of ANSI N44.1-1973 states that non radioactive prototypes may be substituted for radio active units in cases where evidence of integrity can be established by leakage test methods appropriate t(
nonradioactive sources. The integrity of the source will rw established if the source is tested by a method l:'e of accurately and reliably measuring leakage
- ,ies of 10- 7 atm cm3/sec or less for dry air at 25 0C and for a pressure differential of one atmospheric against a vacuum of 1/100 atmosphere or less, with negative results.
2. Item 2 in Section 4.1 states that leakage of less than
50 nanocuries from sources subjected to tests described in Section 4.2 denmonstrates retention of integrity after the test. Prototype sources subjected to tests described in Section 4.2 should retain their integrity and leak essentially no radioactive material. This will be demon strated for the purpose of these prototype tests, if leak tests performed both prior to and after each test described in Section 4.2 indicate removal of less than 5 nanocuries in each case.
3. To the considerations described in Section 5.2,
"Design Requirements," should be added the possible deleterious effects on source materials (both radioactive and nonradioactive) due to the inherent nature of the source and resulting from radiation damage, chemical and physical changes (e.g., those resulting in the evolu tion of gases or changes in density and volume of contained radioactive materials), and chemical reactions between source materials.
4. Section 5.3, "Materials Specification and Control,"
states that the quality control program shall assure that
"substantially the same ....
materials ....
are incorpo rated in all production units." (Emphasis supplied.) All production units should be produced with and contain materials with exactly the same characteristics as the materials used in the qualified prototype units.
5. Section
5.4 of ANSI N44.1-1973,
"Acceptance Testing," states that the acceptability of the source shall be indicated by removal of less than 5 nanocuries of the radioisotope in one of the tests designed to demonstrate contamination of the outer capsule, and it recommends that a value not greater than 1/10 of the radioactivity limit above be chosen as a production control point and that no source which exceeds this value be transferred by a manfacturer to a user for use as a brachytherapy source.
The acceptability of a source for transfer by a manufacturer to a user for use as a brachytherapy source should be indicated by removal of less than 0.5 nanocurie of the radioisotope in one of the tests designed to demonstrate contamination of the outer capsule and by presence of a hermetic seal according to a leakage test designed to demonstrate this situation. The vacuum leach tests, helium mass spectrometer tests, and Kr-85 leak tests 2 are acceptable for the purpose of demonstrating a hermetic seal.
6. Manufacturers of brachytherapy sources should main tain results of prototype and acceptance tests. The test results should be expressed in terms of activity measured and include identification of the test method, instrumen tation, and calibration procedures used.
7. The Atomic Energy Commission does riot license sources and devices containing Ra-226 and its daughters Persons licensed by the AEC for byproduct, source, or special nuclear material who also possess sources con taining radium and its daughters must limit the radiation exposure from both licensed material and other sources of radiation such as radium so that the limits in Part 20
are not exceeded (§20.101). Also, many of the States and all of the Agreement States do regulate the use of sources of radium and its daughters. For that reason, the parts of ANSI N44.1-1973 dealing specifically with testing of sealed radium brachytherapy sources are included in this guide, although the AEC has made no determination as to their adequacy.
2 R.G. Neimeyer,
"Leak Testing Encap.talted Radioactive Sources" ORNL-4529, July
1972, available from National Technical Information Service, U.S. Department of Commerce,
5285 Port Royal Road, Springfield, Va. 22151.
6.2-2