| title = Integrity and Test Specification for Selected Brachytherapy Sources
| title = Rev 1,Integrity and Test Specifications for Selected Brachytherapy Sources
| author name =  
| author name =  
| author affiliation = US Atomic Energy Commission (AEC)
| author affiliation = NRC/RES
| addressee name =  
| addressee name =  
| addressee affiliation =  
| addressee affiliation =  
Line 10: Line 10:
| license number =  
| license number =  
| contact person =  
| contact person =  
| document report number = RG-6.002
| case reference number = RG-6.2 Rev 1
| document type = Regulatory Guide
| document type = Regulatory Guide
| page count = 2
| page count = 2
}}
}}
{{#Wiki_filter:July 1974 U.S. ATOMIC ENERGY COMMISSION
{{#Wiki_filter:U.S. ATOMIC ENERGY COMMISSION  
                                REGULATORY
REGULATORY  
                                DIRECTORATE OF REGULATORY STANDARDS
DIRECTORATE OF REGULATORY STANDARDS
                                                                                                                                GUIDE
Revision 1 July 1974 GUIDE
                                                                  REGULATORY GUIDE 6.2 INTEGRITY AND TEST SPECIFICATIONS FOR SELECTED BRACHYTHERAPY SOURCES
REGULATORY GUIDE 6.2 INTEGRITY AND TEST SPECIFICATIONS FOR SELECTED BRACHYTHERAPY SOURCES


==A. INTRODUCTION==
==A. INTRODUCTION==
sealed medical sources used in interstitial, intracavitary, and topical therapy. This standard was approved by the Manufacture of brachytherapy sources containing                                   American National Standards Institute (ANSI) on byproduct material is an activity requiring a license                                 August 16, 1973, and designated ANSI N44.1-1973. 1 pursuant to section 30.3, "Activities Requiring License,"                             The standard is limited to the traditional type sources of 10 CFR Part 30, "Rules of General Applicability to                                 with well-established medical uses and does not provide Licensing of Byproduct Material;" Brachytherapy                                       standards for sources used in afterloaded devices. The sources manufactured under such license must meet                                     three source type categories covered by the standard are certain integrity requirements and pass certain tests.                                 tubes or capsules and needles containing Ra-266, Co-60,
Manulactuire of brachytherapy sources containing byproduct material is an activity requiring a license pursuant to section 30.3, "Activities Requiring License,"  
Paragraph 32.74(a)(2)(iii) of 10 CFR Part 32, "Specific                               or Cs-137, and beta applicators containing either Sr-90
l' 10) CFR Part 30, "Rules of General Applicability to Licensing of Byproduct Material."  
Licenses to Manufacture, Distribute, or Import Certain                                 or Ra-226. The standard establishes performance test Items Containing Byproduct Material," requires that an                                 specifications for temperature, impact, percussion, bend- application for a specific license to manufacture and                                 ing, tensile stress, and puncture. The temperature, distribute brachytherapy sources and devices containing                               impact, and percussion tests, which are the same tests byproduct material to persons licensed under Section                                   for classification of a source as "special form" under the
Brachytherapy sources manufactured under such license must meet certain integrity requirements and pass certain tests.
35.14, "Special Licenses for Certain Groups of Medical                                 transportation regulations of the Department of Trans- Uses," of 10 CFR Part 35, "Human Uses of Byproduct                                     portation and the International Atomic Energy Agency Material," include a description of procedures for, and                               are applicable to all source types. In addition, the the results of, the prototype tests performed to demon-                                bending test is applicable to interstitial needles, and the strate that the source or device will maintain its integrity                           tensile stress and puncture tests are applicable to the under stresses likely to be encountered in normal use                                 beta applicators. According to ANSI N44.1-1973, a and accidents. Paragraph 32.74(aX2Xiv) of 10 CFR Part                                 source is deemed to pass these tests if it is "free of visual
 
32 requires that the application also include details of                               defects" and "demonstrates leakage of less than 50
Paragraph 32.74(a)(2)(iii) of 10 CFR Part 32, "Specific Licenses to Manufacture, Distribute, or Import Certain Items Containing Byproduct Material," requires that an application for a specific license to manufacture and distribute brachytherapy sources and devices containing byproduct material to persons licensed under Section  
the quality control procedures which assure that pro-                                  nanocuries (nCi) (10 nanocuries of radon in 24 hours in duction sources and devices meet the standards of the                                 the case of Ra-226 sources)."
35.14, "Special Licenses for Certain Groups of Medical Uses," of 10 CFR Part 35, "Human Uses of Byproduct Material," include a description of procedures for, and the results of, the prototype tests performed to demon strate that the source or device will maintain its integrity under stresses likely to be encountered in normal use and accidents. Paragraph 32.74(a)(2)(iv) of 10 CFR Part  
design and prototype tests. This guide describes integrity requirements and test specifications acceptable to the Regulatory staff for selected brachytherapy sources.                                         The standard also includes a quality control program
32 requires that the application also include details of the quality control procedures which assure that pro duction sources and devices meet the standards of the design and prototype tests. This guide describes integrity requirements and test specifications' acceptable to the Regulatory staff for selected brachytherapy sources.
                                                                                      "to insure that the production units will have the same integrity and meet the same requirements as the proto-


==B. DISCUSSION==
==B. DISCUSSION==
type units tested pursuant to this standard." The quality control program further requires that, before transfer, Subcommittee N44-2, Therapeutic Radiology, of the                                each source be tested and demonstrate leakage of less American National Standards Institute Committee on than 5 nanocuries (1 nanocurie of radon in 24 hours in Equipment and Materials for Medical Radiation Appli- the case of Ra-226 sources).
Subcommittee N44-2, Therapeutic Radiology, of the American National Standards Institute Committee on Equipment and Materials for Medical Radiation Appli cation, N44, has developed a standard presenting integ rity requirements and test specifications for selected
cation, N44, has developed a standard presenting integ- rity requirements and test specifications for selected
* Line indicates substantive changes from previous issue.
                                                                                      ]Copies may be obtained from the American" National Standards
 
*Line indicates substantive changes from previous issue.                                Institute, Inc., 1430 Broadway, New York, New York 10018.
sealed medical sources used in interstitial, intracavitary, and topical therapy. This standard was approved by the American National Standards Institute (ANSI)
on August 16, 1973, and designated ANSI N44.1-1973.1 The standard is limited to the traditional type sources with well-established medical uses and does not provide standards for sources used in afterloaded devices. The three source type categories covered by the standard are tubes or capsules and needles containing Ra-266, Co-60,
or Cs-137, and beta applicators containing either Sr-90
or Ra-226. The standard establishes performance test specifications for temperature, impact, percussion, bend ing, tensile stress, and puncture. The temperature, impact, and percussion tests, which are the same tests for classification of a source as "special form" under the transportation regulations of the Department of Trans portation and the International Atomic Energy Agency are applicable to all source types. In addition, the bending test is applicable to interstitial needles, and the tensile stress and puncture tests are applicable to the beta applicators.. According to ANSI N44.1-1973, a source is deemed to pass these tests if it is "free of visual defects" and "demonstrates leakage of less than 50
nanocuries (nCi) (10 nanocuries of radon in 24 hours in the case of Ra-226 sources)."
The standard also includes a quality control program
"to insure that the production units will have the same integrity and meet the same requirements as the proto type units tested pursuant to this standard." The quality control program further requires that, before transfer, each source be tested and demonstrate leakage of less than 5 nanocuries (1 nanocurie of radon in 24 hours in the case of Ra-226 sources).
1Copies may be obtained from the American National Standards Institute, Inc., 1430 Broadway, New York. New York 10018.


USAEC REGULATORY GUIDES                                       Copies of published guides mey be obtained by request indicating the divisions desired to the US. Atomic Energy Commission, Washington, D.C. 20545, Regulatory Guides are issued to describe and make available to the public           Attention: Director of Regulatory Standards. Comments and suggestions for methods acceptable to the AEC Regulatory staff of implementing specific parts of   inmprovements in these guides are encouraged and should be sent to the Secretary the Commission's regulations, to delineate techniques used by the staff in           o1f the Commission, U.S. Atomic Energy Commission, Washington, D.C. 20545.
USAEC REGULATORY GUIDES  
Copies of published guides may be obtained by request indicating the divisions desired to the U.S. Atomic Energy Commission, Washington, D.C. 20545, Regulatory Guides are issued to describe and make available to the public Attention: Director of Regulatory Standards. Comments and suggestions for methods acceptable to the AEC Regulatory staff of implementing specific parts of improvements in these guides are encouraged and should be sent to the Secretary the Commission's regulations, to delineate techniques  
-ed by the staff in of the Commission, U.S. Atomic Energy Commission, Washington, D.C. 20545, evaluating specific problems or postulated accidents, or to provide guidance to Attention: Chief, Public Proceedings Staff.


evaluating specific problems or postulated accidents, or to provide guidance to    Attention: Chief, Public Proceedings Staff.
applicants. Regulatory Guides are not substitutes for regulations and compliance with them is not required. Methods and solutions different from those set out in The guides are issued in the following ten broad divisions:
the guides will be acceptable if they provide a basis for the findings requisite to the issuance or continuance of a permit or license by the Commission.


applicants. Regulatory Guides are not substitutes for regulations and compliance with them is not required. Methods and solutions different from those set out in    The guides are issued in the following ten broad divisions:
===1. Power Reactors ===
  the guides will be acceptable if they provide a basis for the findings requisite to the issuance or continuance of a permit or license by the Commission.                1.  Power Reactors                        
6- Products
2. Research and Test Reactors  


===6. Products===
===7. Transportation ===
                                                                                        2. Research and Test Reactors           
3. Fuels and Materials Facilities
8. Occupational Health Published guides will be revised periodically, as appropriate, to accommodate
4. Environmental and Siting
9. Antitrust Review comments and to reflect new information or experience.


===7. Transportation===
5. Materials and Plant Protection  
                                                                                        3,  Fuels and Materials Facilities        8. Occupational Health Published guides will be revised periodically, as appropriate, to accommodate          4.  Environmental arid Siring              9. Antitrust Review comments and to reflect new information or experience. .                              5. Materials and Plant Protection       1
1


===0. General===
===0. General===


==C. REGULATORY POSITION==
==C. REGULATORY POSITION==
materials with exactly the same characteristics as the iaterials used in the qualified prototype units.
The requirements and recommended practices con tained in ANSI N44. 1-1973, "Integrity and Test Specifi cations for Selected Brachytherapy Sources," 1 con stitute a generally acceptable procedure foi demonstrating the integrity of those medical brachy therapy sources identified in the standard, subject to the following:
1. Section 4.1 of ANSI N44.1-1973 states that non radioactive prototypes may be substituted for radio active units in cases where evidence of integrity can be established by leakage test methods appropriate t(
nonradioactive sources. The integrity of the source will rw established if the source is tested by a method l:'e of accurately and reliably measuring leakage
;,ies of 10- 7 atm cm3/sec or less for dry air at 25 0C and for a pressure differential of one atmospheric against a vacuum of 1/100 atmosphere or less, with negative results.
 
2. Item 2 in Section 4.1 states that leakage of less than
50 nanocuries from sources subjected to tests described in Section 4.2 denmonstrates retention of integrity after the test. Prototype sources subjected to tests described in Section 4.2 should retain their integrity and leak essentially no radioactive material. This will be demon strated for the purpose of these prototype tests, if leak tests performed both prior to and after each test described in Section 4.2 indicate removal of less than 5 nanocuries in each case.
 
3. To the considerations described in Section 5.2,
"Design Requirements," should be added the possible deleterious effects on source materials (both radioactive and nonradioactive) due to the inherent nature of the source and resulting from radiation damage, chemical and physical changes (e.g., those resulting in the evolu tion of gases or changes in density and volume of contained radioactive materials), and chemical reactions between source materials.


,the' requirements and recommended practices con- tained in ANSI N44.1-1973, "Integrity and Test Specifi-            5. Section 5.4 of ANSI N44.1-1973, "Acceptance
4. Section 5.3, "Materials Specification and Control,"  
                                                  1 cations for Selected Brachytherapy Sources,"            con-      Testing," states .that the acceptability of the source shall stitute a generally acceptable procedure for                      be indicated by removal of less than 5 nanocuries of the demonstrating the integrity of those medical brachy-              radioisotope in one of the tests designed to demonstrate therapy sources identified in the standard, subject to the        contamination of the outer capsule, and it recommends following:                                                        that a value not greater than 1/10 of the radioactivity limit above be chosen as a-production control point and
states that the quality control program shall assure that  
1. Section 4.1 of ANSI N44.1-1973 states that non-                that no source which exceeds this value be transferred radioactive prototypes may be substituted for radio-              by a manfacturer to a user for use as a brachytherapy active units in cases where evidence of integrity can be          source.
"substantially the same ....
materials ....  
are incorpo rated in all production units." (Emphasis supplied.) All production units should be produced with and contain materials with exactly the same characteristics as the materials used in the qualified prototype units.


established by leakage test methods appropriate to nonradioactive sources. The integrity of the source will              The acceptability of a source for transfer by a be established if the source is tested by a method                manufacturer to a user for use as a brachytherapy source capable of accuratelT' and reliably measuring leakage            should be indicated by removal of less than 0.5 rates of 107 7 atm cm /sec or less for dry air at 25 0 C and      nanocurie of the radioisotope in one of the tests for a pressure differential of one atmospheric against a          designed to demonstrate contamination of the outer vacuum of 1/100 atmosphere or less, with negative                  capsule and by presence of a hermetic seal according to a results.                                                          leakage test designed to demonstrate, this situation. The vacuum leach tests, helium mass spectrometer tests, and Kr-85 leak tests 2 are acceptable for the purpose of
===5. Section ===
2. Item 2 in Section 4.1 states that leakage of less than          demonstrating a hermetic seal.
5.4 of ANSI N44.1-1973,
"Acceptance Testing," states that the acceptability of the source shall be indicated by removal of less than 5 nanocuries of the radioisotope in one of the tests designed to demonstrate contamination of the outer capsule, and it recommends that a value not greater than 1/10 of the radioactivity limit above be chosen as a production control point and that no source which exceeds this value be transferred by a manfacturer to a user for use as a brachytherapy source.


SO nanocuries from sources subjected to tests described in Section 4.2 demonstrates retention of integrity after          6. Manufacturers of brachytherapy sources should main- the test. Prototype sources subjected to tests described          tain results of prototype and acceptance tests. The test in Section 4.2 should retain their integrity and leak              results should be expressed in terms of activity measured esentially no radioactive materiaL This will be demon-            and include identification of the test method, instrumen- strated for the purpose of these prototype tests, if leak          tatim, and calibration procedures used.
The acceptability of a source for transfer by a manufacturer to a user for use as a brachytherapy source should be indicated by removal of less than 0.5 nanocurie of the radioisotope in one of the tests designed to demonstrate contamination of the outer capsule and by presence of a hermetic seal according to a leakage test designed to demonstrate this situation. The vacuum leach tests, helium mass spectrometer tests, and Kr-85 leak tests 2 are acceptable for the purpose of demonstrating a hermetic seal.


tests performed both. prior to and after each test described in Section 4.2 indicate removal of less than 5            7. The Atomic Energy Commission does not license nanocuries in each case.                                          sources and devices containing Ra-226 and its daughters.
6. Manufacturers of brachytherapy sources should main tain results of prototype and acceptance tests. The test results should be expressed in terms of activity measured and include identification of the test method, instrumen tation, and calibration procedures used.


Persons licensed by the AEC for byproduct, source, or
7. The Atomic Energy Commission does riot license sources and devices containing Ra-226 and its daughters Persons licensed by the AEC for byproduct, source, or special nuclear material who also possess sources con taining radium and its daughters must limit the radiation exposure from both licensed material and other sources of radiation such as radium so that the limits in Part 20  
3. To the comiderations described in Section 5.2,                  special nuclear material who also possess sources con-
are not exceeded (§20.101). Also, many of the States and all of the Agreement States do regulate the use of sources of radium and its daughters. For that reason, the parts of ANSI N44.1-1973 dealing specifically with testing of sealed radium brachytherapy sources are included in this guide, although the AEC has made no determination as to their adequacy.
"Design Requirements," should be added the possible                taining radium and its daughters must limit the radiation deleterious effects on source materials (both radioactive          exposu-e from both licensed material and other sources and nonradioactive) due to the inherent nature of the              of radiation such as radium so that the limits in Part 20
source and resulting from radiation damage, chemical                are not exceeded (§20.101). Also, many of the States and physical changes (e.g., those resulting in the evolu-          and all of the Agreement States do regulate the use of tion of gases or changes in density and volume of                  sources of radium and its daughters. For that reason, the contained radioactive materials), and chemical reactions          parts of ANSI N44.1-1973 dealing specifically with between source materials.                                          testing of sealed radium brachytherapy sources are included in this guide, although the AEC has made no determination as to their adequacy.


4. Section 5.3, "Materials Specification. and Control,"
2 R.G. Neimeyer,  
states that the quality control program shall assure that          2 R.G. Neimeyer, "Leak Testing. Encapsulated Radioactive
"Leak Testing Encap.talted Radioactive Sources" ORNL-4529, July  
"substantially the same .... materials .... are incorpo-            Sources" ORNL-4529, July 1972, available from National rated in all production units." (Emphasis supplied.) All            Technical Information Service, U.S. Department of Commerce, production units should be produced with and contain                5285 Port Royal Road, Springfield, Va. 22151.
1972, available from National Technical Information Service, U.S. Department of Commerce,  
5285 Port Royal Road, Springfield, Va. 22151.


6.2-2}}
6.2-2}}


{{RG-Nav}}
{{RG-Nav}}

Latest revision as of 02:10, 17 January 2025

Rev 1,Integrity and Test Specifications for Selected Brachytherapy Sources
ML003739394
Person / Time
Issue date: 07/31/1974
From:
Office of Nuclear Regulatory Research
To:
References
RG-6.2 Rev 1
Download: ML003739394 (2)


U.S. ATOMIC ENERGY COMMISSION

REGULATORY

DIRECTORATE OF REGULATORY STANDARDS

Revision 1 July 1974 GUIDE

REGULATORY GUIDE 6.2 INTEGRITY AND TEST SPECIFICATIONS FOR SELECTED BRACHYTHERAPY SOURCES

A. INTRODUCTION

Manulactuire of brachytherapy sources containing byproduct material is an activity requiring a license pursuant to section 30.3, "Activities Requiring License,"

l' 10) CFR Part 30, "Rules of General Applicability to Licensing of Byproduct Material."

Brachytherapy sources manufactured under such license must meet certain integrity requirements and pass certain tests.

Paragraph 32.74(a)(2)(iii) of 10 CFR Part 32, "Specific Licenses to Manufacture, Distribute, or Import Certain Items Containing Byproduct Material," requires that an application for a specific license to manufacture and distribute brachytherapy sources and devices containing byproduct material to persons licensed under Section

35.14, "Special Licenses for Certain Groups of Medical Uses," of 10 CFR Part 35, "Human Uses of Byproduct Material," include a description of procedures for, and the results of, the prototype tests performed to demon strate that the source or device will maintain its integrity under stresses likely to be encountered in normal use and accidents. Paragraph 32.74(a)(2)(iv) of 10 CFR Part

32 requires that the application also include details of the quality control procedures which assure that pro duction sources and devices meet the standards of the design and prototype tests. This guide describes integrity requirements and test specifications' acceptable to the Regulatory staff for selected brachytherapy sources.

B. DISCUSSION

Subcommittee N44-2, Therapeutic Radiology, of the American National Standards Institute Committee on Equipment and Materials for Medical Radiation Appli cation, N44, has developed a standard presenting integ rity requirements and test specifications for selected

  • Line indicates substantive changes from previous issue.

sealed medical sources used in interstitial, intracavitary, and topical therapy. This standard was approved by the American National Standards Institute (ANSI)

on August 16, 1973, and designated ANSI N44.1-1973.1 The standard is limited to the traditional type sources with well-established medical uses and does not provide standards for sources used in afterloaded devices. The three source type categories covered by the standard are tubes or capsules and needles containing Ra-266, Co-60,

or Cs-137, and beta applicators containing either Sr-90

or Ra-226. The standard establishes performance test specifications for temperature, impact, percussion, bend ing, tensile stress, and puncture. The temperature, impact, and percussion tests, which are the same tests for classification of a source as "special form" under the transportation regulations of the Department of Trans portation and the International Atomic Energy Agency are applicable to all source types. In addition, the bending test is applicable to interstitial needles, and the tensile stress and puncture tests are applicable to the beta applicators.. According to ANSI N44.1-1973, a source is deemed to pass these tests if it is "free of visual defects" and "demonstrates leakage of less than 50

nanocuries (nCi) (10 nanocuries of radon in 24 hours2.777778e-4 days <br />0.00667 hours <br />3.968254e-5 weeks <br />9.132e-6 months <br /> in the case of Ra-226 sources)."

The standard also includes a quality control program

"to insure that the production units will have the same integrity and meet the same requirements as the proto type units tested pursuant to this standard." The quality control program further requires that, before transfer, each source be tested and demonstrate leakage of less than 5 nanocuries (1 nanocurie of radon in 24 hours2.777778e-4 days <br />0.00667 hours <br />3.968254e-5 weeks <br />9.132e-6 months <br /> in the case of Ra-226 sources).

1Copies may be obtained from the American National Standards Institute, Inc., 1430 Broadway, New York. New York 10018.

USAEC REGULATORY GUIDES

Copies of published guides may be obtained by request indicating the divisions desired to the U.S. Atomic Energy Commission, Washington, D.C. 20545, Regulatory Guides are issued to describe and make available to the public Attention: Director of Regulatory Standards. Comments and suggestions for methods acceptable to the AEC Regulatory staff of implementing specific parts of improvements in these guides are encouraged and should be sent to the Secretary the Commission's regulations, to delineate techniques

-ed by the staff in of the Commission, U.S. Atomic Energy Commission, Washington, D.C. 20545, evaluating specific problems or postulated accidents, or to provide guidance to Attention: Chief, Public Proceedings Staff.

applicants. Regulatory Guides are not substitutes for regulations and compliance with them is not required. Methods and solutions different from those set out in The guides are issued in the following ten broad divisions:

the guides will be acceptable if they provide a basis for the findings requisite to the issuance or continuance of a permit or license by the Commission.

1. Power Reactors

6- Products

2. Research and Test Reactors

7. Transportation

3. Fuels and Materials Facilities

8. Occupational Health Published guides will be revised periodically, as appropriate, to accommodate

4. Environmental and Siting

9. Antitrust Review comments and to reflect new information or experience.

5. Materials and Plant Protection

1

0. General

C. REGULATORY POSITION

The requirements and recommended practices con tained in ANSI N44. 1-1973, "Integrity and Test Specifi cations for Selected Brachytherapy Sources," 1 con stitute a generally acceptable procedure foi demonstrating the integrity of those medical brachy therapy sources identified in the standard, subject to the following:

1. Section 4.1 of ANSI N44.1-1973 states that non radioactive prototypes may be substituted for radio active units in cases where evidence of integrity can be established by leakage test methods appropriate t(

nonradioactive sources. The integrity of the source will rw established if the source is tested by a method l:'e of accurately and reliably measuring leakage

,ies of 10- 7 atm cm3/sec or less for dry air at 25 0C and for a pressure differential of one atmospheric against a vacuum of 1/100 atmosphere or less, with negative results.

2. Item 2 in Section 4.1 states that leakage of less than

50 nanocuries from sources subjected to tests described in Section 4.2 denmonstrates retention of integrity after the test. Prototype sources subjected to tests described in Section 4.2 should retain their integrity and leak essentially no radioactive material. This will be demon strated for the purpose of these prototype tests, if leak tests performed both prior to and after each test described in Section 4.2 indicate removal of less than 5 nanocuries in each case.

3. To the considerations described in Section 5.2,

"Design Requirements," should be added the possible deleterious effects on source materials (both radioactive and nonradioactive) due to the inherent nature of the source and resulting from radiation damage, chemical and physical changes (e.g., those resulting in the evolu tion of gases or changes in density and volume of contained radioactive materials), and chemical reactions between source materials.

4. Section 5.3, "Materials Specification and Control,"

states that the quality control program shall assure that

"substantially the same ....

materials ....

are incorpo rated in all production units." (Emphasis supplied.) All production units should be produced with and contain materials with exactly the same characteristics as the materials used in the qualified prototype units.

5. Section

5.4 of ANSI N44.1-1973,

"Acceptance Testing," states that the acceptability of the source shall be indicated by removal of less than 5 nanocuries of the radioisotope in one of the tests designed to demonstrate contamination of the outer capsule, and it recommends that a value not greater than 1/10 of the radioactivity limit above be chosen as a production control point and that no source which exceeds this value be transferred by a manfacturer to a user for use as a brachytherapy source.

The acceptability of a source for transfer by a manufacturer to a user for use as a brachytherapy source should be indicated by removal of less than 0.5 nanocurie of the radioisotope in one of the tests designed to demonstrate contamination of the outer capsule and by presence of a hermetic seal according to a leakage test designed to demonstrate this situation. The vacuum leach tests, helium mass spectrometer tests, and Kr-85 leak tests 2 are acceptable for the purpose of demonstrating a hermetic seal.

6. Manufacturers of brachytherapy sources should main tain results of prototype and acceptance tests. The test results should be expressed in terms of activity measured and include identification of the test method, instrumen tation, and calibration procedures used.

7. The Atomic Energy Commission does riot license sources and devices containing Ra-226 and its daughters Persons licensed by the AEC for byproduct, source, or special nuclear material who also possess sources con taining radium and its daughters must limit the radiation exposure from both licensed material and other sources of radiation such as radium so that the limits in Part 20

are not exceeded (§20.101). Also, many of the States and all of the Agreement States do regulate the use of sources of radium and its daughters. For that reason, the parts of ANSI N44.1-1973 dealing specifically with testing of sealed radium brachytherapy sources are included in this guide, although the AEC has made no determination as to their adequacy.

2 R.G. Neimeyer,

"Leak Testing Encap.talted Radioactive Sources" ORNL-4529, July

1972, available from National Technical Information Service, U.S. Department of Commerce,

5285 Port Royal Road, Springfield, Va. 22151.

6.2-2