Attachment 3 - Passport Leak Test 092719
ML19290E327
Person / Time
Site: 07007022
Issue date: 09/27/2019
From: Korbly S, Casey Smith
F. X. Masse Associates, Passport Systems
To: Ty Naquin
NRC/NMSS/DFM/FFLB
Naquin T
References
Download: ML19290E327 (1)


Text

Institution: Passport Systems Date:

9/27/2019 Locations: 70 Treble Cove Rd Billerica MA Date wipes were taken:

8/12/2019 Count Time:

1 min Background Count Rate (A):

46 cpm Background Count Rate (B):

25 cpm Radioisotope Swipe #

Initial Activity Manufacturer SN PSI Designator Net CPM A,B Instrument Efficiency A,B Measured activity of Leak Test (µCi)

HEU 1

T3E67-0001 N/A 12, 0 A,B 1.4, >90%

< 1 x 10-3 HEU Disk 2

T3E67-0002 N/A 3, 2 A,B 1.4, >90%

< 1 x 10-3 HEU Disk 3

T3E67-0003 N/A 23, 1 A,B 1.4, >90%

< 1 x 10-3 LEU Plate 4

Plate 6 N/A 3, 2 A,B 1.4, >90%

< 1 x 10-3 LEU Plate 5

Plate 8 N/A 22, 4 A,B 1.4, >90%

< 1 x 10-3 Leak Test Wipes Analyzed By:

Chad A. Smith PhD, CHP

  • Wipes were analyzed using a (A) Health Physics Instruments Model 7010 Rainbow MCA (SN 701188) with a Health Physics Instruments Model 5530/7530 NaI Scillation Probe (SN 730288) and (B)

PerkinElmer Tri-Carb 2910 TR liquid scintillation counter.

Semi-Annual Leak Test Results F.X. MASSÉ ASSOCIATES, INC.

Health and Medical Physics Consultants MA Registration 65-0016 NH Registration 258 RI Registration RPS0015 24 Hours: 978-283-4888 PO Box 100 Gloucester, MA 01931-0100 TEL 978-283-4888 FAX 978-281-6702 info@fxmasse.com