ML19211C147

From kanterella
Jump to navigation Jump to search
Shine Medical Technologies, LLC, Quality Assurance Program Description
ML19211C147
Person / Time
Site: SHINE Medical Technologies
Issue date: 07/17/2019
From:
SHINE Medical Technologies
To:
Office of Nuclear Reactor Regulation
References
2019-SMT-0054
Download: ML19211C147 (26)


Text

25 pages follow ENCLOSURE 14 SHINE MEDICAL TECHNOLOGIES, LLC SHINE MEDICAL TECHNOLOGIES, LLC APPLICATION FOR AN OPERATING LICENSE QUALITY ASSURANCE PROGRAM DESCRIPTION

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 3 of 27 TABLE OF CONTENTS Executive Summary...................................................................................................................... 5 Policy Statement........................................................................................................................... 5 1

Introduction............................................................................................................................ 6 1.1 Scope............................................................................................................................. 6 1.2 Application...................................................................................................................... 6 1.3 Definitions....................................................................................................................... 7 2

Design, Construction, and Modifications............................................................................... 8 2.1 Organization................................................................................................................... 8 2.2 Quality Assurance Program.......................................................................................... 11 2.3 Design Control.............................................................................................................. 12 2.4 Procurement Document Control................................................................................... 14 2.5 Procedures, Instructions and Drawings........................................................................ 14 2.6 Document Control......................................................................................................... 14 2.7 Control of Purchased Items and Services.................................................................... 15 2.8 Identification and Control of Items................................................................................ 16 2.9 Control of Special Processes....................................................................................... 16 2.10 Inspections................................................................................................................... 16 2.11 Test Control.................................................................................................................. 17 2.12 Control of Measuring and Test Equipment................................................................... 17 2.13 Handling, Storage and Shipping................................................................................... 17 2.14 Inspection, Test and Operating Status......................................................................... 17 2.15 Control of Non-conforming Items and Services............................................................ 17 2.16 Corrective Actions........................................................................................................ 18 2.17 Quality Records............................................................................................................ 18 2.18 Assessments................................................................................................................ 18 3

Facility Operations............................................................................................................... 19 3.1 Organization................................................................................................................. 19 3.2 Quality Assurance Program.......................................................................................... 19 3.3 Performance Monitoring............................................................................................... 19

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 4 of 27 3.4 Operator Experience.................................................................................................... 19 3.5 Operating Conditions.................................................................................................... 20 3.6 Operational Authority.................................................................................................... 20 3.7 Control Area................................................................................................................. 20 3.8 Ancillary Duties............................................................................................................. 20 3.9 Emergency Communications........................................................................................ 20 3.10 Configuration Control.................................................................................................... 20 3.11 Lockouts and Tagouts.................................................................................................. 21 3.12 Test and Inspection...................................................................................................... 21 3.13 Operating Procedures.................................................................................................. 21 3.14 Operator Aid Postings.................................................................................................. 21 3.15 Equipment Labeling...................................................................................................... 21 4

Applicability to Existing Facilities......................................................................................... 21 5

Decommissioning................................................................................................................ 22 6

References.......................................................................................................................... 22 - SHINE Functional Organizational Chart.............................................................. 23 - Graded Approach to Quality................................................................................ 24 - Procedures that Impliment the QAPD.................................................................. 25

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 5 of 27 EXECUTIVE

SUMMARY

This Quality Assurance Program Description (QAPD) provides the Shine Medical Technologies, LLC (SHINE) quality assurance program (QAP) for safe and reliable production of 99Mo and other radioisotopes for medical use and is specific to SHINE. Title 10 of the Code of Federal Regulations (CFR), § 50.34(a)(7) requires each applicant for a construction permit to build a production or utilization facility to include, in its Final Safety Analysis Report (FSAR), a description of the QAP to be applied to the design, fabrication, construction, and testing of the structures, systems, and components (SSCs) of the facility. Furthermore, 10 CFR § 50.34(b)(6)(ii) requires that each applicant for a license to operate a facility include, in the FSAR, a description of the managerial and administrative controls to be used to assure safe operation.

NUREG-1537, Guidelines for Preparing and Reviewing Applications for the Licensing of Non-Power Reactors, Section 12.9, Quality Assurance, recommends the applicant consider the guidance in Regulatory Guide 2.5, Quality Assurance Program Requirements for Research and Test Reactors, and ANSI/ANS 15.8, Quality Assurance Program Requirements for Research Reactors. Regulatory Guide 2.5, Rev.1 states that ANSI/ANS 15.8-1995 provides an acceptable method of complying with the program requirements of 10 CFR § 50.34 and was used for developing this QAPD. SHINE has determined that ANSI/ANS 15.8-1995 is sufficient for use in the development of the SHINE QAPD, which is to be applied to the design, fabrication, construction, and operation of the SHINE facility.

POLICY STATEMENT SHINE shall design, procure, operate and maintain the SHINE facility in a manner that will ensure the health and safety of the public and workers and protect the environment. SHINE commits to operating in a planned, prudent, efficient, safe and cost-effective manner that is in compliance with the CFR, the applicable Nuclear Regulatory Commission (NRC) Facility Operating License, and the applicable laws and regulations of the state and local governments.

In addition, the management of SHINE believes that sound quality, safety, security, and environmental programs are essential to SHINEs success and is personally engaged in their implementation. Quality is owned by every SHINE employee. Quality cannot be achieved solely through inspection. It is achieved by employees with the knowledge, skills, experience, training and motivation to do the job right. Persons who manage, perform and verify work all contribute to an integrated, cost effective and efficient quality management system to produce a quality product.

The SHINE QAP is the QAPD provided in this document and the associated implementing documents. They provide control over SHINE activities that affect the quality of safety-related nuclear plant SSCs and include all planned and systematic activities necessary to provide adequate confidence that such SSCs will perform satisfactorily in service. The QAPD is the top-level document that establishes the manner in which quality is to be achieved and presents the SHINE overall philosophy regarding the achievement and assurance of quality. Implementing documents assign more detailed responsibilities and requirements and define organizational interfaces involved in conducting activities within the scope of the QAP. SHINE uses the guidance

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 6 of 27 and best practices from various industry standards to develop implementing procedures. SHINE applies a graded approach consistent with importance to safety and reliability.

Compliance with the QAPD and implementing documents is mandatory for personnel directly or indirectly associated with implementation of the SHINE QAPD.

1 INTRODUCTION SHINE uses a new class of isotope generator that is compact and relatively inexpensive to generate a reliable supply of Mo-99 and other radioisotopes for medical applications. This technology does not use highly enriched uranium, nor does it require a nuclear reactor for production.

Although SHINE uses an accelerator for production, this technology requires licensing under 10 CFR Part 50, Domestic Licensing of Production and Utilization Facilities. This QAPD addresses the requirements in 10 CFR § 50.34(b)(6)(ii) for a description of a QAP specific for SHINE. Regulatory Guide 2.5 and ANSI/ANS 15.8-1995 provide an acceptable method of complying with the quality assurance program requirements of 10 CFR § 50.34.

SHINE uses accelerator technology for neutron production and the sub-critical fission process and does not meet the definition of a nuclear reactor as identified in 10 CFR § 50.2, Definitions.

Nonetheless, SHINE uses a definition of safety-related for systems, structures and components (SSCs) that is applied to the Quality Level 1 SSCs and uses a graded approach to quality for other SSCs. The graded approach to quality for this QAPD can be found in Enclosure 2.

1.1 Scope SHINE addresses the requirements of 10 CFR § 50.34 for a description of the QAP in this controlled document. This QAPD and applicable implementing procedures apply specifically to SHINE. The procedures that implement the requirements in this document are found in the SHINE Information Management System.

The QAPD describes the administrative and engineered controls for ensuring compliance with requirements. It applies to the design, construction and operation of the SHINE facility.

1.2 Application The quality assurance program applied by SHINE activities will be consistent with the importance of these activities to safety and reliability. Activities included in this quality assurance program are, as a minimum, those related to irradiation unit safety and protection system, material processing safety, criticality safety, engineered safety features, and applicable radiation monitoring systems, as identified in the Limiting Conditions for Operations sections of the Technical Specifications.

SHINE applies a graded approach to those items and activities that could affect the quality of safety-related structures, systems and components (SSCs) and other components not specifically designated as safety-related. A Quality Level (QL) matrix is used to ensure quality requirements are understood and specified for each SSC. Activities that could affect quality include siting,

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 7 of 27 designing, purchasing, fabricating, handling, shipping, receiving, storing, cleaning, erecting, installing, repairing, maintaining, modifying, inspecting, testing, and operating.

SHINE developed and implemented the QAP beginning with the design, siting and construction phase of the facility. This QAP focuses on the development of appropriate controls to ensure the facility is properly designed and fabricated to meet SHINE requirements. The majority of these controls provide documentation that attests to the facility quality to support an operating license.

Following facility construction and commissioning, the focus of this quality program shifts to establishing those controls that ensure proper and reliable facility operation. All of the program provisions established during the design and construction phase will remain in place, but will change in level of implementation appropriate to support facility operations; each portion of Section 2, Design, Construction and Modifications, would be implemented only as necessary. The operating phase will impose additional requirements related to the conduct of operations.

Additional program requirements are defined in Section 3.

1.3 Definitions Definitions are listed to provide uniform interpretation of terms and phrases used.

certified operator - an individual authorized by the chartering or licensing organization to carry out the duties and responsibilities associated with the position requiring the certification.

commissioning - the process during which constructed irradiation facility structures, components, and systems are made operational and verified to meet design requirements.

corrective action - measures taken to rectify conditions adverse to quality, and where necessary, to prevent repetition.

document - any written or pictorial information describing, defining, specifying, reporting, or certifying activities, requirements, procedures, or results.

licensed operator - see certified operator.

maintenance - those activities necessary to maintain operability or restore systems to within specified design limits. Maintenance consists of repair, rework, replacement, adjustment, cleaning, or other actions necessary to maintain an item in or restore an item to an acceptable condition.

management - management means those persons within the SHINE organization whose responsibility and authority includes the quality assurance program. The levels of management are as described in ANSI/ANS-15.1-2007.

modification - a change in the physical design or functional characteristics of a system, structure, or component.

procedure - a document that specifies or describes how an activity is to be performed.

quality - the degree to which an item or process meets or exceeds the users requirements and expectations.

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 8 of 27 quality assurance - those planned and systematic actions necessary to provide adequate confidence that the structure, system, or component will perform satisfactorily in service.

safety-related activities - those activities affecting the safety-related functions of SSCs, including siting, designing, purchasing, fabricating, handling, shipping, receiving, storing, cleaning, erecting, installing, repairing, maintaining, modifying, inspecting, testing, and operating.

safety-related Items - Those physical SSCs whose intended functions are to prevent accidents that could cause undue risk to health and safety of workers and the public; and to control or mitigate the consequences of such accidents.

shall, should and may - the word "shall" is used to denote a requirement; the word "should" to denote a recommendation; and the word "may" to denote permission, neither a requirement nor a recommendation.

2 DESIGN, CONSTRUCTION, AND MODIFICATIONS This section provides the requirements for establishing, managing, conducting, and assessing the program of controls over the design, construction, and modification of the SHINE facility. SHINE recognizes that the described controls are integral to the management of the licensed activity and do not necessitate the establishment of a separate program. This section will be implemented as applicable to the specific scope of work activities.

2.1 Organization This section describes the SHINE organizational structure, functional responsibilities, levels of authority and interfaces for establishing, executing, and verifying QAPD implementation. The organizational structure includes internal and external functions for SHINE including interface responsibilities for multiple organizations that perform quality-related functions. The organizational structure and assignment of responsibilities are defined and documented such that: (a) quality is achieved and maintained by those who have been assigned responsibility for performing work; and (b) quality achievement is verified by those not directly performing the work. SHINE Senior Management ensures t h a t t h o s e responsible for ensuring appropriate controls have been established, and for verifying that activities have been correctly performed, h a v e sufficient authority, access to work areas, and freedom to: (a) identify problems; (b) initiate, recommend, or provide corrective action; and (c) ensure corrective action implementation. Implementing documents assign more specific responsibilities and duties, and define the organizational interfaces involved in conducting activities and duties within the scope of the QAPD. Management gives careful consideration to the timing, extent and effects of organizational structure changes. The SHINE functional organization chart is provided in Enclosure 1.

Chief Executive Officer (CEO)

The CEO is responsible for the overall design, management and technical leadership of the company and is also responsible for all technical and administrative support activities provided by SHINE and suppliers. The CEO provides direction to the President & Chief Financial Officer, Chief Operating Officer (COO), Vice President of Regulatory Affairs and Quality (VPRA/Q),

(CFO), and Information Technology Manager (ITM) to fulfill the organizations responsibilities. The CEO reports to the Board of Directors with respect to all matters.

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 9 of 27 The CEO has overall responsibility for SHINE QA Program.

President & Chief Financial Officer (P&CFO)

The P&CFO is second-in-command at SHINE and acts for the CEO when the CEO is not in the office or otherwise not available. The P&CFO reports to the CEO and is responsible for all financial matters for SHINE. The P&CFO will oversee all compliance and recognition for government (federal and state) contracts and private grants. The P&CFO is also responsible for procurement and all external operations of SHINE, including supplier organizations.

Procurement Manager (PrM)

The PrM reports to the P&CFO and is responsible for overall company policy regarding the procurement of material, equipment, and services; establishes specific procedures for the contracting of services and the purchase and control of materials and equipment. The PrM is also responsible for integrating and implementing applicable requirements from the QAPD into the procurement process to ensure that suppliers meet SHINE requirements.

Additionally, the PrM is responsible for the oversight of suppliers and the management aspects associated with their execution of the design, fabrication, procurement, construction and operation of the SHINE facility.

Supplier Organizations Supplier organizations are responsible for their portion of the execution of the design, fabrication, procurement, construction, and testing of SSCs for the facility. Such supplier organizations are responsible for identifying, implementing and verifying flow-down of quality requirements as applicable. They will participate in necessary assessments and inspections as specified in procurement documents.

Chief Technology Officer (CTO)

CTO reports to the CEO and is responsible for expediting the path to commercial production through identification and reduction of technical risks, simplifying design requirements, and driving the development of innovative solutions.

Chief Operating Officer (COO)

The COO reports to the CEO and is responsible for Engineering and the operational aspects of the company including safety, management, and training. The COO is also responsible for the development and implementation of the SHINE Document Control and Records Management programs as well as for matters regarding environment, safety, and health. The COO works closely with the Director of Quality (DoQ) on matters involving adherence to safety requirements defined in the QAPD and other regulatory, state and local requirements.

Director of Process Engineering (DoE)

The DoE is responsible for nuclear, instrumentation and control and chemical process systems and system testing. The DoE is also responsible for maintaining the safety analysis and is the design authority for the facility. The DoE reports to the COO.

Director of Engineering Construction (DoC)

The DoC is responsible all aspects of facility construction, including modification activities, project management and field engineering. structural design, auxiliary equipment

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 10 of 27 (mechanical, electrical and plumbing), construction support and field engineering. The DoC reports to the COO.

Director of Engineering Support (DoS)

The DoS is responsible for facility design configuration control and engineering support.

The DoS is also responsible for the development and implementation of the SHINE Document Control and Records Management program. The DoS reports to the COO.

Plant Manager (PM)

The PM is responsible for the operation and management of SHINEs facilities. The PM is also responsible for establishing and managing the required training programs to support the organization, and for establishing and maintaining the programs and systems to ensure protection of the companys assets. The PM reports to the COO.

Operations Manager (OM)

The OM is responsible for safe, reliable, and efficient plant operations within the constraints of the operating license and regulatory requirements. This position is also responsible for the development and implementation of appropriate operational controls in accordance with the QAPD and other requirements. The OM reports to the PM.

Radiation Protection Manager (RPM)

The RPM is responsible for establishing and implementing the Radiation Protection (RP) program, monitoring worker doses, and calibration and QA of all health physics instrumentation. The RPM reports to the PM.

Training Manager (TM)

The TM is responsible for development, administration and overall management of the implementation of the various training processes and activities, including as required by regulations and the training for site and facility personnel. The TM reports to the PM.

Vice President of Regulatory Affairs and Quality (VPRA/Q)

The VPRA/Q reports to the CEO and is responsible for nuclear related licensing and quality activities. The VPRA/Q is responsible for the planning and execution of the licensing process for the design, construction and operation of the facility. The VPRA/Q is also responsible for the implementation of the SHINE Corrective Action program. The VPRA/Q is responsible for ensuring clear lines of communication between SHINE and the NRC. The VPRA/Q is responsible for integrating the quality requirements as defined in the QAPD across the internal and external organizations and reports to the CEO on matters concerning quality. The VPRA/Q is also responsible for all local, state and federal permits.

Director of Quality (DoQ)

The DoQ is responsible for Quality Assurance and Quality Control (QA/QC) processes and activities, including the development and verification of implementation of the QAPD described in this document. The DoQ is responsible for assuring compliance to regulatory requirements and procedures through assessments and technical reviews; monitoring organizational processes to ensure conformance to commitments and licensing document requirements; for ensuring that contractors and suppliers providing quality services, equipment, parts, and materials to SHINE are meeting the requirements as defined in the

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 11 of 27 QAPD. The DoQ has sufficient independence from other priorities to bring forward issues affecting safety and quality and makes judgments regarding quality in all areas necessary regarding SHINEs activities. The DoQ has the ability and responsibility to report to the CEO any quality issues which cannot be resolved at the VPRA/Q level. The DoQ reports to the VPRA/Q.

Director of Licensing (DoL)

The DoL is responsible for implementing nuclear related licensing activities, including the planning and execution of the licensing process for the design, construction and operation of the facility. The DoL is responsible for communication between SHINE and the NRC.

The DoL reports to the VPRA/Q.

Information Technology Manager (ITM)

The ITM provides technical expertise regarding electronic data systems, computer networks, and telecommunications and ensuring the technology necessary for implementation of the QAPD and implementing procedures is available. The ITM reports to the CEO.

Authority to Stop Work All employees have the right and responsibility to stop work when they encounter an unsafe condition. Additionally, quality assurance and inspection personnel have the authority and the responsibility to stop work in progress which is not being done in accordance with approved procedures or where safety-related SSC quality may be jeopardized. This extends to off-site work performed by suppliers that furnish materials and services.

Quality Assurance Organizational Independence Independence shall be maintained between the organizations performing the work or service and oversight performed by the quality organization (i.e., quality assurance and quality control).

2.2 Quality Assurance Program This section describes the requirements for establishing, implementing, and managing the QAP for SHINE in accordance with the requirements in ANSI/ANS-15.8-1995, Quality Assurance Program Requirements for Research Reactors. This includes the managerial and administrative aspects of internal and external activities that affect quality of the SHINE facility and programs.

To achieve the goals of defining and effectively designing safety-related SSCs, SHINE implements the use of a graded approach to quality. The measures applied to a particular engineered or administrative control or control system may be graded commensurate with the reduction of the risk attributable to that control or control system. This approach to achieving levels of quality is described in this QAPD and related implementing documents.

The QAPD provides the basis for a planned and systematic approach to the cost-effective achievement of safety, quality and reliability. The primary method to ensure this is through the SHINE procedures. The SHINE procedures are delineated, managed and maintained by the VPRA/Q with support from all SHINE team members. See Enclosure 3 for a matrix of QAPD sections and their corresponding implementing procedures.

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 12 of 27 Delegated responsibilities may be performed under a contractors or suppliers QAP, provided that they have been approved in accordance with the QAPD. Periodic assessments of their QAP are performed to ensure compliance with the SHINE QAPD and implementing procedures. In addition, routine interfaces with their personnel provide added assurance that quality expectations are met. Assessments may be planned and performed by SHINE qualified assessors or independent contractors or consultants as determined by the QM.

Personnel assigned to implement elements of the QAPD shall be capable of performing their assigned tasks. SHINE establishes and maintains formal and informal indoctrination and training programs for personnel performing, verifying or managing activities within the scope of the QAPD to ensure that suitable proficiency is achieved and maintained. Sufficient managerial depth is provided to cover absences of incumbents. When required by code, regulation or standard, specific qualification and selection of personnel is conducted in accordance with those requirements as established in applicable SHINE procedures. Indoctrination includes the administrative and technical objectives and requirements of the applicable codes and standards and the QAPD requirements as necessary. Records of personnel training and qualification are to be maintained.

2.3 Design Control This section describes the requirements for establishing, developing, implementing and documenting a process to control the design, design changes, and temporary modifications subject to the provisions of the QAPD. Procedures identify the process and include provisions for the control of design, development, verification, approval, release, status, distribution, revisions, review of calculations and other design documents, control of software and implementation of required rules, regulations, codes and standards. As part of the design control, the design review program has been developed to meet the requirements of ANSI/ANS-15.8-1995.

Design Requirements Applicable design inputs, such as design bases, performance requirements, regulatory requirements, codes, and standards shall be identified and documented.

Design Process Design interfaces shall be identified and controlled and the design efforts shall be coordinated by the design organization among the participating organizations. Interface controls will include the assignment of responsibility and establishment of implementing documents among the interfacing design organizations for the review, approval, release, distribution and revision of documents involving design interfaces.

The applicability of standardized or previously proven designs, with respect to meeting pertinent design inputs, shall be verified for each application. Known problems affecting the standardized or previously proven designs, and their effects on other features, shall be considered. Deviations from the established and documented design inputs, including the reasons for the changes, shall be documented and controlled.

The design organization is responsible to ensure that the final design shall:

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 13 of 27

1. be relatable to design input by documentation in sufficient detail to permit design traceability and verification, and
2. identify assemblies and/or components that are part of the item being designed.

When a computer design program is used to develop portions of the facility design or to analyze a design for acceptability, that program shall be fully documented, validated and controlled to ensure the correctness of its output. When a design program must be developed, the program shall be controlled to ensure that it is fully documented and validated. Where changes to previously valid computer programs are made, documented revalidation shall be required for the change. Verification of design-unique computer programs shall include appropriate benchmark testing.

Design Verification Independent design reviews shall be used to verify the adequacy of design by one or more of the following:

1. performance design reviews,
2. use of alternate calculations,
3. performance of qualification tests, or
4. comparison of similar proven systems.

The responsible design organization shall identify and document the particular design verification method or methods used. Design verification will be performed by competent individuals or groups other than those who performed the design, but whom may be from the same organization. In all cases the design verification shall be completed prior to reliance upon the component, system, structure, or computer program to perform its function in operations.

In the event that qualification testing is needed to verify design, the use of qualification tests will be defined in a formal test plan that shall include appropriate acceptance criteria and shall demonstrate the adequacy of performance under conditions that simulate the most adverse design conditions. Test results will be documented and evaluated by the responsible design organization to ensure that test requirements have been met.

Design Documents and Records Design documents and records, which provide evidence that the design and design verification process were performed, shall be collected, stored and maintained for the life of the safety-related unit.

Commercial Grade Items The use of commercial-grade equipment in safety-related applications shall be reviewed to ensure that it can adequately perform its intended function. Procedures shall be implemented to provide guidance on how to review and evaluate commercial grade items for suitability in applications covered by the QAPD. When a commercial grade item, prior to its installation, is modified or selected by special inspection and/or testing to requirements that are more restrictive than the suppliers published product description,

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 14 of 27 the component part shall be represented as different from the commercial grade item in a manner traceable to a documented definition of the difference.

Change Control Procedures shall be established to ensure that modifications to safety-related structures, systems, and components, or computer codes shall be based on a defined as-exists design. Changes to verified designs shall be documented, justified, and subject to design control measures commensurate with those applied to the original design. The control measures shall include assurance that the design analyses for the structure, system, component, or computer code are still valid. Where a significant design change is necessary because of an incorrect design, the design process and verification procedure shall be reviewed and modified as necessary.

2.4 Procurement Document Control Procedures shall be established to ensure that procurement documents will contain sufficient technical and quality requirements to ensure that the items or services satisfy the needs of SHINE.

Procurement documents at all procurement levels shall identify the documentation required to be submitted for information, review, or approval by SHINE. At each level of procurement, the procurement documents shall provide for access to the suppliers plant facilities and records, for inspection or assessment by SHINE, a designated representative or other parties authorized by SHINE.

SHINE procurement documents shall include SHINEs requirements for reporting and approving disposition of suppliers non-conformances associated with the items or services being procured.

The procurement documents for safety-related items should prohibit the supply of sub-standard or counterfeit parts or materials.

2.5 Procedures, Instructions and Drawings Activities affecting quality shall be performed in accordance with documented instructions, procedures, or drawings appropriate to the circumstances. These documents shall include or reference appropriate quantitative or qualitative acceptance criteria for determining that activities have been satisfactorily accomplished.

These documents shall be prepared to prescribe performance expectations and define the proper sequence and detail to accomplish the work. Copies of applicable and necessary procedures, instructions and drawings shall be available to the appropriate SHINE internal and external organizations to accomplish work in an efficient and safe manner. See Enclosure 3 for a detailed listing of implementing procedures.

2.6 Document Control The preparation, issue, and change of documents which specify requirements that affect quality or prescribe activities affecting quality, shall be controlled to ensure that correct documents are used. The document control system shall be documented, and provide for:

1. identification of documents to be controlled and their specified distribution;

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 15 of 27

2. identification of assignment of responsibility for preparing, reviewing, approving, and issuing documents; and
3. review of documents for adequacy, completeness, and correctness prior to approval and issuance.

Major changes to controlled documents shall be reviewed and approved by the same organizations that performed the original review and approval, unless other organizations are specifically designated.

2.7 Control of Purchased Items and Services The procurement of items and services shall be controlled to ensure appropriate procurement planning, source evaluation and selection, evaluation of objective evidence of quality furnished by the supplier, source inspection, assessment and examination of items or services for acceptance upon delivery or completion.

Supplier Selection The selection of suppliers shall be based on evaluation of their capability to provide items or services in accordance with requirements of the procurement documents.

Work Control SHINE shall establish measures to control the suppliers performance to ensure that purchased items and services meet quality requirements. Controls may include test plans, review of suppliers submitted documents, arrangements for source surveillance or inspection, and other technical and administrative interfaces with the supplier in accordance with procurement documents.

Verification Activities The supplier shall be responsible for the quality of his product and shall verify and provide evidence of that quality. Supplier-generated documents shall be controlled, handled, and approved in accordance with established methods. Means shall be implemented to provide for the acquisition, processing, and recorded evaluation of technical, inspection and test data against acceptance criteria. Based on complexity of the product and importance to safety, SHINE shall consider independently verifying the quality of a suppliers product through source surveillances, inspections, assessments or review of the suppliers non-conformances, dispositions, waivers and corrective actions.

Item or Service Acceptance Acceptance of items or services provided to SHINE shall require a system to provide assurances that purchased items and services conform to procurement specifications.

Methods used to accept an item or related service from a supplier shall be a supplier Certificate of Conformance, source verification, receiving inspection, post-installation test or a combination thereof. Receiving inspection shall be performed in accordance with established procedures and instructions, to verify by objective evidence such features as proper configuration, identification and cleanliness, and to determine any shipping damage, fraud or counterfeit.

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 16 of 27 2.8 Identification and Control of Items When specified by codes, standards, or specifications that include identification or traceability requirements, the item identification and control process shall be capable of providing identification and traceability control. Items identification shall be maintained from the initial receipt or fabrication of the items up to and including installation and use. Where physical identification on the item is either impractical or insufficient, physical separation, procedural control, or other appropriate means shall be employed. Identification markings shall be applied through the use of materials and methods which provide clear and legible identification and do not detrimentally affect the function or service life of the item. Markings shall be transferred to each part of an identified item when the item is subdivided and shall not be obliterated or hidden by surface treatment or coatings unless substitute means are provided. Where specified, items having limited calendar or operating life shall be identified and controlled to preclude use of items whose shelf life or operating life is expired.

2.9 Control of Special Processes Special processes include any in which the results are highly dependent on the control of the process or skill of the personnel. These are also those processes in which the specified quality cannot be readily determined by inspection or non-destructive testing of the product. Special processes shall be controlled by instructions, procedures, drawings, checklists, travelers or other appropriate means. SHINE and its suppliers are responsible for adhering to the approved procedures and processes for performing the special process. The requirements of applicable codes and standards, including acceptance criteria for the process, shall be specified or referenced in the procedures or instructions that control the process. Records shall be maintained as appropriate for the currently qualified personnel, processes, and equipment associated with special processes.

2.10 Inspections Inspections to verify conformance of an item or activity to requirements shall be planned, documented and performed. The inspection program shall apply to procurement, construction, modification, and maintenance. Inspection of items in-process or under construction shall be performed for work activities where product quality cannot be determined by inspection of the completed product. The final inspection shall be planned to arrive at a conclusion regarding conformance of the item to specified requirements. Completed items shall be inspected for completeness, markings, calibration, adjustments, protection from damage or other characteristics as required to verify the quality and conformance of the item to specified requirements. Associated quality records shall be examined for adequacy and completeness.

Only items that have passed the required inspections and tests shall be used, installed or operated. Measuring and Test Equipment (M&TE) used to perform inspections shall be identified in inspection documentation for traceability of inspection results.

Inspection results shall be documented. Acceptance of items shall be documented and approved by authorized personnel. Inspection shall be performed by persons other than those who performed the work being inspected, but may be from the same organization. Each person who verifies conformance of work activities for purposes of acceptance shall be qualified to perform the assigned inspection task. The need for formal training shall be determined and training

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 17 of 27 activities conducted as required to qualify personnel who perform inspections and tests. On-the-job training shall be included, with emphasis on firsthand experience gained through actual performance of inspections. Records of inspection personnels qualification shall be established and maintained by their employer.

2.11 Test Control Formal testing shall be required to verify conformance of designated structures, systems or components to specified requirements and demonstrate satisfactory performance for service or to collect data in support of design or fabrication. Testing shall include prototype qualification tests, proof tests prior to installation and functional tests. Test results shall be documented and evaluated by a responsible authority to ensure that test requirements have been satisfied.

Computer programs used for operational control shall be tested in accordance with an approved verification and validation plan and shall demonstrate required performance over the range of operation of the controlled function or process.

2.12 Control of Measuring and Test Equipment Tools, gauges, instruments and other M&TE used for activities affecting quality shall be controlled and calibrated or adjusted, at specified periods to maintain accuracy within specified limits. Out-of-calibration devices shall be tagged or segregated, and not used until they have been recalibrated. Records shall be maintained of calibration data traceable to the individual piece of M&TE. Calibration and control measures are not required when normal commercial equipment provides adequate accuracy.

2.13 Handling, Storage and Shipping Handling, storage, and shipping of items shall be in accordance with work and inspection instructions, drawings, specifications, shipping instructions or other pertinent documents or procedures for conducting the activity.

2.14 Inspection, Test and Operating Status The status of inspection and test activities shall be identified on the items or in documents traceable to the items, in order to ensure that required inspections and tests are performed and to ensure that items which have not passed the required inspections and tests are not inadvertently installed or operated.

2.15 Control of Non-Conforming Items and Services Items that do not conform to requirements shall be controlled to prevent inadvertent installation or use. Controls on non-conforming items shall provide for identification, documentation, evaluation, segregation from like conforming items when practical, and disposition of non-conforming items. Non-conforming conditions shall be evaluated for further reporting to appropriate regulatory agencies. Non-conforming characteristics shall be reviewed, and recommended dispositions of non-conforming items proposed and approved, in accordance with documented procedures.

The disposition (use-as-is, reject, repair, or rework) of non-conforming items shall be identified and documented. Technical justification for the acceptability of a non-conforming item disposition

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 18 of 27 repair or use-as-is shall be documented. Non-conformance to design requirements of items dispositioned use-as-is or repair shall be subject to design control measures commensurate with those applied to the original design. The as-built records shall reflect the accepted deviation.

Repaired or reworked items shall be re-examined in accordance with applicable procedures and with the original acceptance criteria, unless the non-conforming item disposition has established alternate acceptance criteria.

2.16 Corrective Actions Conditions adverse to quality shall be identified promptly and corrected as soon as practical. The corrective actions shall be in accordance with the design requirements unless those requirements were faulty. In the case of a significant condition adverse to quality, the cause of the condition shall be investigated and corrective action taken to preclude recurrence.

2.17 Quality Records A records system or systems shall be established at the earliest practical time consistent with the schedule for accomplishing work activities. The records system or systems shall be defined, implemented, and enforced in accordance with written procedures, instructions, or other documentation. The records shall include as a minimum: inspection and test results, results of quality assurance reviews, quality assurance procedures, and engineering reviews and analyses in support of designs or changes and modifications.

Some records shall be maintained by or for the plant owner for the life of the particular item while it is installed in the plant or stored for future use. Such records shall be classified in accordance with the following criteria:

1. those which would be of value in demonstrating capability for safe operation;
2. those which would be of value in maintaining, reworking, repairing, replacing, or modifying an item;
3. those which would be of value in determining the cause or results of an accident or malfunction of a safety-related item;
4. those which provide required baseline data for in-service inspections; or
5. those which would be of value in planning for facility decommissioning.

Other records shall be retained for a shorter period as determined by SHINE. The records shall be stored in a location or locations that prevent damage from moisture, temperature, and pestilence. Additional provisions shall be made for special processed records such as radiographs, photographs, negatives, microfilm and magnetic media, to prevent damage from excessive light, stacking, electromagnetic fields, temperature and humidity. Records maintained by a supplier shall be accessible to SHINE.

2.18 Assessments SHINE will conduct periodic assessments of quality-affecting activities during design, construction, modification and operations to evaluate the effectiveness of the as-implemented quality program. Assessments shall be performed in accordance with written procedures or checklists. Assessment results shall be documented, and should be reviewed by management personnel who have responsibility for the area assessed. Conditions requiring prompt corrective

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 19 of 27 action shall be reported immediately to the appropriate management of the assessed organization.

Management of the assessed organization or activity shall investigate adverse findings, schedule corrective action (including measures to prevent recurrence) and notify the appropriate assessing organization in writing of action taken or planned. The adequacy of the responses shall be evaluated by the assessing organization. Assessment records include assessment plans, reports, written replies, and the record of completion of corrective action. Personnel selected for assessment assignments shall have experience or training commensurate with the scope, complexity, or special nature of the activities to be assessed. The assessor shall have the capability to communicate effectively, both in writing and orally.

3 FACILITY OPERATIONS This section provides the elements of the QAP for conduct of operation at the SHINE facility. The requirements shall be applied to any equipment or operation as appropriate and consistent with its potential safety impact or program goals. Many of the program requirements are satisfied by existing documentation, or by procedures and activities required by other standards and requirements of the chartering or licensing agency. Some requirements of the QAP for operations may also be found in other documents, such as the Training Program, Emergency Plan, Security Plan, Technical Specifications (see ANSI/ANS-15.1-2007) and the Radiation Protection Program (see American National Standard for Radiation Protection at Research Reactor Facilities, ANSI/ANS-15.11-2009). Such requirements do not need to be duplicated in the quality assurance program.

3.1 Organization SHINE shall provide sufficient resources in personnel and materials to safely conduct operations.

Planning should anticipate needs as appropriate for any task. The organization structure shall be defined as required by Technical Specifications.

3.2 Quality Assurance Program SHINE shall establish a QAP by implementing a policy for the conduct of operations. The policy will assign personnel to implement the policy and identify the goals for operating the SHINE facility. Personnel assignments and progress toward achieving goals will be documented.

3.3 Performance Monitoring SHINE shall monitor facility performance relative to the goals that will be used as performance indicators. SHINE shall document periodic observations of operations and identify and assess any deficiencies to ensure the execution of corrective actions that will prevent recurrence. If appropriate, trend analysis should be performed to indicate where improvements or lessons learned could be implemented. Violations of operating practices should be addressed and documented as appropriate.

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 20 of 27 3.4 Operator Experience SHINE shall document the methods for maintaining operator experience. Operators should be responsible for maintaining experience in operating the SHINE facility. This may be achieved by routine operation of the SHINE facility and documentation of the activity. A method should be provided to make operators aware of important current information that is related to facility operations and individual job assignments. Operator training is addressed in American National Standard for the Selection and Training of Personnel for Research

Reactors, ANSI/ANS 15.4-2016.

3Property "ANSI code" (as page type) with input value "ANSI/ANS 15.4-2016.</br></br>3" contains invalid characters or is incomplete and therefore can cause unexpected results during a query or annotation process..5 Operating Conditions Pre-operations checklists shall be used to determine or verify required pre-operational conditions and readiness to operate. Operating equipment shall be periodically monitored to detect abnormal conditions or adverse trends. Operating conditions should be documented in an operations logbook or other record. The operator should notify the appropriate level of management of any abnormal situations.

3.6 Operational Authority SHINE shall establish the method for conducting operations and the responsibility for each shift.

Operating personnel shall conduct a comprehensive review of appropriate records and equipment before assuming responsibility for the facility. Operational authority may be transferred through a documented turnover briefing and facility walk-through procedures. These procedures should include checklists to record items important to facility status.

3.7 Control Area Operators shall be alert and attentive to the control consoles indications, alarms, and other activities within the control area. Only persons specifically authorized or certified to operate the SHINE facility shall operate control area equipment. Trainees may operate equipment only when they are directly supervised by certified operators. Control area activities and access should be limited to ensure that the operators are attentive to control responsibilities. A procedure shall be in place for quick placement of the SHINE facility in a safe configuration if evacuation of the control area or site is necessary.

3.8 Ancillary Duties Operators shall not be assigned ancillary duties to be performed during operations to the extent that these duties could interfere with the ability to monitor facility parameters and maintain control of the SHINE facility.

3.9 Emergency Communications Operators shall be able to contact the appropriate level of management rapidly and shall have the means to notify all affected personnel promptly of operations or emergencies on-site.

3.10 Configuration Control Equipment shall be identified that requires configuration control. SHINE is responsible for establishing and maintaining proper configuration and should authorize any changes to safety-related items. All configuration changes to safety-related items should be documented. Before

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 21 of 27 placing equipment into operation, the system shall be properly calibrated or checked, as appropriate, and any deficiencies in the equipment or the current configuration of the system documented. This should also address methods for temporary modifications. SHINE facility maintenance that requires a change in the system shall be documented.

3.11 Lockouts and Tagouts Locks and tags shall be placed on equipment when, for safety or other special administrative reasons, controls must be established. If there is potential for equipment damage or personnel injury during equipment operation, maintenance, inspection, or modification activities, or from inadvertent activation of equipment, a facility lockout/tagout procedure shall be implemented.

3.12 Test and Inspection Tests shall be performed following system maintenance, design changes, or inspection that involves dismantlement of components or systems. A documented test plan shall be used to demonstrate that the component or system is capable of performing its intended function. The results of the test should be documented and retained in facility records as appropriate.

3.13 Operating Procedures Operating procedures shall provide appropriate direction to ensure that the facility is operated normally within its design basis, and in compliance with technical specifications. Operating procedures shall be written, reviewed, approved by appropriate management, controlled, and monitored to ensure that the content is technically correct and the wording and format are clear and concise. The facility policy on use of procedures should be documented and clearly understood by all operators. The extent of detail in a procedure should depend on the complexity of the task; the experience, education, and training of the users; and the potential significance of the consequences of error. The process for making changes and revisions to procedures should be documented. A controlled copy of all operations procedures should be maintained in the control room or equivalent area.

3.14 Operator Aid Postings Any posted information that aids operators in performing their duties should be current and correct. Management should review operator aids to determine that they are necessary and correct before approving their postings. Postings should be checked periodically for continued applicability.

3.15 Equipment Labeling Equipment shall be labeled to help facility personnel positively identify equipment they operate and maintain. Information on labels should be consistent with information found in facility procedures, valve lineup sheets, piping and instrument diagrams or other documents. Labels should be permanent, securely attached, readable and have appropriate information.

4 APPLICABILITY TO EXISTING FACILITIES The SHINE facility will be a newly constructed facility and this section does not apply.

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 22 of 27 5 DECOMMISSIONING This section of the QAPD will be updated at a later date.

6 REFERENCES 6.1 American National Standard for Quality Assurance Program Requirements for Research Reactors, ANSI/ANS-15.8-1995.

6Property "ANSI code" (as page type) with input value "ANSI/ANS-15.8-1995.</br></br>6" contains invalid characters or is incomplete and therefore can cause unexpected results during a query or annotation process..2 American National Standard for the Development of Technical Specifications for Research Reactors, ANSI/ANS-15.1-2007.

6Property "ANSI code" (as page type) with input value "ANSI/ANS-15.1-2007.</br></br>6" contains invalid characters or is incomplete and therefore can cause unexpected results during a query or annotation process..3 American National Standard for the Selection and Training of Personnel for Research Reactors, ANSI/ANS-15.4-2016.

6Property "ANSI code" (as page type) with input value "ANSI/ANS-15.4-2016.</br></br>6" contains invalid characters or is incomplete and therefore can cause unexpected results during a query or annotation process..4 Appendix B to 10 CFR Part 20Annual Limits on Intake (ALIs) and Derived Air Concentrations (DACs) of Radionuclides for Occupational Exposure; Effluent Concentrations; Concentrations for Release to Sewerage.

6.5 Final Interim Staff Guidance Augmenting NUREG 1537, Part 1 Guidelines for Preparing and Reviewing Applications for the Licensing of Non-Power Reactors: Format and Content, for licensing Radioisotope Production Facilities and Aqueous Homogeneous Reactors, October 17, 2012.

6.6 NUREG 1537, Part 1 Guidelines for Preparing and Reviewing Applications for the Licensing of Non-Power Reactors, Format and Content.

6.7 Regulatory Guide 2.5 Rev.1, Quality Assurance Requirements for Research and Test Reactors.

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 23 of 27 ENCLOSURE 1 - SHINE FUNCTIONAL ORGANIZATIONAL CHART

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 24 of 27 ENCLOSURE 2 - GRADED APPROACH TO QUALITY The graded approach to quality is a process by which the level of analysis, documentation, and actions necessary to comply with a requirement is commensurate with the safety significance.

The graded approach permits the implementing organization to focus resources on those activities that are deemed, by qualitative analysis, to reduce the associated risks and hazards.

The activities and tasks are performed in accordance with approved implementing procedures.

QL-1 shall implement the full measure of this QAPD and shall be applied to safety-related SSCs and to safety-related activities.

QL-2 is applied to SSCs and activities that are intended to prevent unacceptable interactions between non-seismic Category I SSCs and safety-related SSCs.

QL-3 is applied to nonsafety-related SSCs and activities. SSCs and activities may either be designated as commercial or portions of this QAPD may be applied. The performance of QL-3 SSCs and activities may be important to ensuring customer requirements are met, or operational or mission-related goals such as throughput, cost, or schedule are achieved.

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 25 of 27 ENCLOSURE 3 - PROCEDURES THAT IMPLEMENT THE QAPD QAPD Section Subject SHINE Implementing Procedures Baker Implementing Procedures 2.1 Organization QAPD, section 2.1 01.01 Organization 2.1.1 Authority to Stop Work 0100-01-01 Stop Work Baker QAPD, section 5.1.3E 2.1.2 QA Organizational Independence QAPD, section 2.1.2 01.01 Organization 2.2 QA Program QAPD, section 2.2 2900-01-01 Training 1200-01-07 Classification of SSCs 02.01 General Co-worker Indoctrination & Training 02.02 Co-workers Qualification and Certification 02.03 Qualtiy Assurance Program Description Procedure 2.3 2.3.1 2.3.2 2.3.3 2.3.4 Design Control Design Requirements Design Process Design Verification Design Documents and Records 1200-01-02 Calculations 1200-01-03 Owners Acceptance Review 1200-01-08 Drawings 1200-01-10 Design Control 1200-01-15 Specifications for SSCs 1200-09-01 Configuration Management 1200-01-13 Software Control and QA 03.01 Design Interface 2.3.5 Commercial Grade Items 1600-01-02 Commercial Grade Dedication 07.02 Commercial Grade Dedication (CGD) of Items for Safety Related Applications 2.3.6 Change Control 1200-01-06 Design Change Control Process 2.4 Procurement Document Control 1600-01-01 Procurement 04.01 Controlling Procurement Documents 2.5 Procedures, Instructions and Drawings 1200-01-08 Drawings 2100-01-01 Document Control 5.01 Controlling Instructions and Procedures 2.6 Document Control 2100-01-01 Document Control 2000-01-03 Documentation Practices 2000-01-12 Signatures and Review/Approval Process 06.01 Controlling Documents

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 26 of 27 QAPD Section Subject SHINE Implementing Procedures Baker Implementing Procedures 2.7 2.7.1 2.7.2 2.7.3 2.7.4 Control of Purchased Items and Services Supplier Selection Work Control Verification Activities Item or Service Acceptance 2000-01-11 Supplier Qualification 1600-01-01 Procurement 1600-01-03 Request for Quotation (RFQ) 2000-01-13 QA Surveillances Receipt Inspection procedure to be developed for facility operations prior to OL issuance.

Baker procedure used for construction.

07.01 Controlling Purchases of Items and Services 2.8 Identification and Control of Items Procedure to be developed for facility operations prior to OL issuance. Baker procedure used for construction.

08.01 Identifying and Marking Material 2.9 Control of Special Processes Procedure to be developed for facility prior to OL issuance.

Baker procedure used for construction.

09.01 Controlling Special Processes 2.10 Inspection Procedure to be developed for facility prior to OL issuance.

Baker procedure used for construction.

10.01 Performing Inspections 2.11 Test Control 1200-01-13 Software Control and QA Procedure for non-software testing to be developed for facility prior to OL issuance. Baker procedure used for construction.

11.01 Controlling Tests 2.12 Control of Measuring and Test Equipment Procedure to be developed for facility prior to OL issuance.

Baker procedure used for construction.

12.01 Controlling Measuring and Test Equipment 2.13 Handling, Storage and Shipping Procedure to be developed for facility prior to OL issuance.

Baker procedure used for construction.

13.01 Handling, Storage, Cleaning, Packaging and Shipping of Items 2.14 Inspection, test and Operating Status Procedure to be developed for facility operations prior to OL issuance. Baker procedure used for construction.

14.01 Identifying the Inspection, Test and Operating Status of Items 2.15 Control of Nonconforming Items and Services 2000-01-14 Control of Nonconforming Items.

15.01 Controlling Nonconforming Items

2000-09-01 Rev. 13 Quality Assurance Program Description SHINE Program Page 27 of 27 QAPD Section Subject SHINE Implementing Procedures Baker Implementing Procedures 2.16 Corrective Actions 2200-01-01 Issues Management 2200-01-03 10 CFR 21 and 10 CFR § 50.55(e) Reporting 2200-01-05 Condition Evaluation 2200-01-06 Apparent Cause Evaluation 2200-01-07 Root Cause Evaluation 16.01 Corrective Action Program 16.02 Reporting Potential 10 CFR 21 & 10 CFR § 50.55(e) Issues to the SHINE Issue Management Committee Technical Lead 2.17 Quality Records 2100-01-02 Records Management 2000-11-01 SHINE Quality Records Retention Policy 17.01 Controlling Quality Assurance Records 2.18 Assessments 2000-01-08 Assessments 2000-01-13 QA Surveillances 18.01 Performing Assessments 18.02 Performing Surveillances 3

Facility Operations Procedures to be developed for facility operations prior to OL issuance.

Baker scope for construction only